17
4.3 Interface Section
Displayed when [7. I.F.
] is set to [1] or [2]
Item
Set Value
Description
Output Control
71. o.c.
0 Stop output
1 Output continuous at all times
2 Output continuous if stable (stop output if unstable)
3
Outputs once by pressing Print key (irrespective of
whether stable).
4
Outputs once if stable. Outputs if the balance is
stable when a sample is loaded after the preceding
sample has been removed and the balance indicated
zero, or less.
5
Outputs once if stable, and stops output when
unstable. Even if the sample is not replaced, the
balance is output once when it stabilizes next time
(including the zero indication).
6
Outputs once if stable, and outputs continuously when
unstable. Even if the sample is not replaced, output
of the balance stops when it stabilizes after being
output once.
☆
7
Pressing Print key causes the balance to output
once when stable.
Baud Rate
72. b.L.
☆
1 1200 bps
2 2400 bps
3 4800 bps
4 9600 bps
Parity
73. PA.
☆
0 None
Displayed only when [7. I.F. 2 ]
(7-digit numeric format) is specified.
1 Odd
2 Even
☆
denotes a factory-setting
※
The data interval in continuous output mode is 0.1 to 1 second. (The interval varies depending
on weighting conditions and other factors.)