Calibrate Video

Calibrate

Calibrate process

Calibration is the process to ensure the device will scan with the optimal 
accuracy and scan quality. When the software installed the first time, it 
automatically goes to calibration interface.

Different  calibration  boards  are  used  for  scanning  ranges  of  300mm 
and  150mm.  Select  the  corresponding  calibration  board  as  shown  in 
the calibration interface.

Note:

1.Make  sure  to  protect  the  calibration  board  and  keep  it  clean,  no 
scratches or stains on both side. 

2.The  calibration  board  is  matched  to  the  Device  with  same  Serial 
Number. Doing the calibration with an incorrect calibration board will fail 
to generate good scan data or optimum accuracy. 

3.Clean with pure water only, do not use alcohol or other chemical liquid 
to clean the calibration board. 

4.To prevent damage to the calibration board, do not drop the board, 
and do not place heavy objects or irrelevant objects on the board. 

5.After usage, store the calibration board in the velvet bag immediately.

Calibrate

Cameras calibrate

White Balance 

Calibration

Finish Calibration

Operate positions according 

to the software wizard

Start to capture data

Capturing completed

Calibration succeed 

Calibration failed

Restart equipment 

adjustment

Summary of Contents for Transcan C

Page 1: ...A Professional 3D Scanner for Diverse Industries Getting Started with Transcan C User Manual Transcan C ...

Page 2: ...M 4G USB port high speed USB 3 0 port 1 USB 2 0 port Power adapter 2 Power cable 2 Calibration board 150mm 300mm Calibration board bracket Device connection cable Calibration pad Marker Turntable USB drive Turntable connection cable Tripods Scanner head Light Box Recommendation Power 60W lumen 12000 13000LM input voltage 110 240V color temperature 5500K 200K ...

Page 3: ...ve of the tripod adjust the orientation and tighten the screws to fix it as shown 5 Based on the need shake the rocker to adjust the device height Then tighten the latch Hardware Installation Scanner Adjustment 1 Confirm that power switch is not pressed 2 Connect the power cable to the adapter port first 3 Inserted the adapter socket into the device port 4 Plugin the power adapter to power source ...

Page 4: ...nning depends on whether open the lightbox while doing the white balance test 5 Make sure the light box connection cable is connected in the correct order and the ports connected to each lamp are connected to one to four adapter cable Lightbox connection optional 1 Connect the scanner lightbox cable to the lightbox power cable 2 Connect the scanner lightbox cable to the one to four connection cabl...

Page 5: ...allation Equipment Adjustment Software Download STEP 01 STEP 02 1 Software installation 2 Software activation 3 Scanner adjustment 4 Select scanning range 5 Adjust camera position according to the range 6 Adjust projector focus 7 Adjust camera angle 8 Adjust camera aperture 9 Adjust camera focus 10 Turntable lightbox connection check ...

Page 6: ...n board is matched to the Device with same Serial Number Doing the calibration with an incorrect calibration board will fail to generate good scan data or optimum accuracy 3 Clean with pure water only do not use alcohol or other chemical liquid to clean the calibration board 4 To prevent damage to the calibration board do not drop the board and do not place heavy objects or irrelevant objects on t...

Page 7: ...ing Data measurement Data reservation Scan preparation Create Import project Check working distance Select turntable mode brightness and resolution Texture scan Non texture vscan Scan Edit Scanning Data Scan completed Alignment Continue to scan ...

Page 8: ...Strongly surface reflective objects Shiny and black object Objects that undergo deformation Hollow objects such as Eiffel Tower souvenirs Hair and similar lint like structures Spray on the surface Solution Recommend not to scan Scan technics ...

Page 9: ...Summarize Scan Range mm Accuracy mm 0 05 300 X 190 Point Distance mm Alignment Mode 0 03 0 07 0 11 0 06 0 15 0 23 150 X 96 Marker Alignment Feature Alignment Manually Alignment ...

Page 10: ... China P 86 571 82999050 Email sales shining3d com No 1398 Xiangbin Road Wenyan Xiaoshan Hangzhou Zhejiang China 311258 Americas Region SHINING 3D Technology Inc San Francisco United States P 1415 259 4787 Email sales shining3d com 1740 César Chávez St Unit D San Francisco CA 94124 EMEA Region SHINING 3D Technology GmbH Stuttgart Germany P 49 711 28444089 Email sales shining3d com Breitwiesenstraß...

Reviews: