Calibrate Video
Calibrate
Calibrate process
Calibration is the process to ensure the device will scan with the optimal
accuracy and scan quality. When the software installed the first time, it
automatically goes to calibration interface.
Different calibration boards are used for scanning ranges of 300mm
and 150mm. Select the corresponding calibration board as shown in
the calibration interface.
Note:
1.Make sure to protect the calibration board and keep it clean, no
scratches or stains on both side.
2.The calibration board is matched to the Device with same Serial
Number. Doing the calibration with an incorrect calibration board will fail
to generate good scan data or optimum accuracy.
3.Clean with pure water only, do not use alcohol or other chemical liquid
to clean the calibration board.
4.To prevent damage to the calibration board, do not drop the board,
and do not place heavy objects or irrelevant objects on the board.
5.After usage, store the calibration board in the velvet bag immediately.
Calibrate
Cameras calibrate
White Balance
Calibration
Finish Calibration
Operate positions according
to the software wizard
Start to capture data
Capturing completed
Calibration succeed
Calibration failed
Restart equipment
adjustment