Calibration
3.1.
Precautions and Use
Calibration is the process to ensure the device will scan with the optimal accuracy
and scan quality. The parameters of the scanner will be recalculated during the
calibration.
Under the following situations, you should calibrate the device:
◆
Device change
◆
After device enduring bumpy transportation
◆
After device accuracy decreases
◆
The device has not been calibrated for 15 days or more
◆
After device plugged in with color camera
⚠Notes:
Make sure to protect the calibration board and keep it clean, no scratches or
stains on the black surface with white circles.
The Calibration board is matched to the Device with same Serial Number. Doing
the calibration with an incorrect calibration board will fail to generate good scan
data or optimum accuracy.
Clean with clear water only, do not use alcohol or chemical liquid to clean the
calibration board
3.2.
Operation
After installation, when you open the software for the first time the software will
enter the calibration mode by default. You can also choose Calibration on the
navigation bar to enter calibration process later.