User’s Manual for Palm Optical Fiber Tester
Shineway Technologies, Inc.
- 5 -
2. Basic Operation
2.1Foreword
This part introduces the basic operations of the ShinewayTech
series hand-held
optical testing instruments. Specific operations of each type are elaborated in “Specific
Operation”, chapter 4 of this manual. Please read this manual carefully for optimal operation.
Should you encounter any problems during operation, you are welcome to contact the
technical staff of our company or agents
2.2
Battery Installation and Replacement
ShinewayTech
series palm optical instruments use an alkaline battery. Before use of
the instrument, make sure the battery is correctly installed.
When the battery power is low, low battery indicator is shown on LCD. As long as
there is any character on LCD, the instrument can still be in operation. However, when LCD
turns dark, the output power becomes unstable, therefore measured power value will not be
accurate and in consequence, the measurement accuracy is degraded.
2.3
Turn off Auto-off Function
With default setting, the Auto-off function of ShinewayTech
series palm optical fiber
tester is on, for this can conserve power and battery life. Auto-off function needs to be
turned off, when long-time test is performed. Please follow the following procedures.
Turn off Auto-off:
VLS: when power on the instrument, hold
[On/Off]
until the indicator light turns
green to disable Auto-off.
SLS, OPM and OLT
:
when power on the instrument, hold
[On/Off]
until letter “P”
is displayed on LCD to disable Auto-off. In the coming measurement, “AUTO OFF” will
not be displayed on LCD. With the default setting, the series will conserve battery life by
automatically turning itself off if no keys have been pressed for approximately 5 minutes.
With default setting, the Auto-off function is on after each power on.
2.4
Cleaning of Connectors
Please observe the following safety instructions when cleaning this instrument: