52
11. Environment
Read Chapter 7 Before Use of
These Pages
11.3.2
Timing of Stability Mark Illumination and Data Output
The illumination of stability mark and automatic data output of the functions employing stability detec-
tion take place at the same timing. The timing after detection of stability is optimized at the default and
it is not necessary to change the setting usually.
However, adjustment of the timing is allowed to three levels according to specific measurement
requirements. When wishing to accelerate automatic data output with functions such as animal weigh-
ing, select menu item
. When wishing to delay the illumination of stability mark, select menu item
. Select menu item
to return it to the standard setting.
11.4 Tracking
Tracking is the function that will maintain the current displayed value as long as possible.
To turn ON this function,
Select menu item
.
To turn OFF this function,
Select menu item
.
Zero Tracking Function
When the display is zero, tracking functions as “zero tracking” to keep the
zero display as long as possible. “Zero tracking” automatically cancels
small zero drift.
It is recommended to set tracking off when measuring slight mass change
such as in the process of drop addition or liquid evaporation.
Note
Using a verified balance as a legal measuring instrument in the EU:
Even when Tracking is ON, the range where it works is restricted to the
vicinity of the zero, and it is subjected to conditions, according to the regu-
lations.