76
-
Note:
is displayed during the process. Leave the balance in a stable
state until the mass display appears as in step 6.
•
In PCAL, the value of the "weight to be loaded" cannot be changed.
•
Set the PCAL password using Menu Map item [71].
2.7 Graphic
Display
Menu
:
2.7.1 Overview
The BX-K and BW-K series balance has a bar graph located on the left side of the display.
This graph can be conveniently used for liquid weighing and pass or failure judgment
(comparator function) of the sample.
The graphic display functions include the Full Scale mode, Target mode, Group mode, Level
mode, and no graphic display mode.
Note:
Target, limit, upper, and lower values are set as numeric values only. Set the correct
numeric value for the unit that will be used for weighing.
•
In weighing mode, changing the displayed unit does NOT change the target, limit,
upper, or lower numeric values. For example, when the upper limit has been set at
10g, switching the unit from “g” to “kg” does not change the limit to 10kg. It will
be 0.01kg.
•
The decimal point is invisible in the value setting display. Determine its position
based on the resolution of the balance. For example, if the balance’s minimum
display is 0.01g, 150g must be set by inputting “15000” in the display.
•
Numeric values are memorized independently for each graphic display mode.
•
The graphic display mode can be selected without setting numeric values.
2.7.2 Full Scale Mode
The relative amount of the load on the pan is displayed in the bar graph. This feature helps
to prevent errors due to OL (overload) status.
Use Menu Map item [11] to select the Full Scale mode.
A bar displayed in the lower areas of the scale indicates that the load on the pan is small.
A bar displayed in the upper areas of the scale indicates that the load on the pan is close to
the weighing capacity.
2.7.3 Target Mode
Target mode is useful for filling applications or for quality check by weight.
The
value is the numeric value that is the desired amount in the unit that is
used for weighing. The
value is the numeric amount above or below the target
value that is acceptable.
This mode is useful for constant amount weighing of liquid or judgment of excess and
shortage.
Menu Map item [12]
selects the Target mode.
Summary of Contents for B252KS
Page 2: ......