145
9
9
Connections and Communications with External Equipment
Connections and Communications with External Equipment
1
3
5
2
4
6
7
8
12
14
13
9
10
11
GLP output Function
It is possible to add the balance ID, user name and other information as the header when outputting
records of calibration on weighing values by setting the GLP output Function at ON. A signature field
can also be included as a footer.
The header and the footer are not output even when the GLP function is set at ON if the interval timer
function is being used.
If the software versions is 1.01.28 and later, and 1.01.I.01 and later, when setting the GLP output
Function at ON, header and footer are added only to records of calibration. The header and the
footer are not output when outputting weighing values.
Refer to the following for details on how to verify the software version.
^
[Software Version Verification Method] (P.26)
1.
Call out the calibration menu
Press [
CAL] for three or more consecutive
seconds.
The calibration menu will be displayed.
2.
Set GLP output at ON or OFF
(1) Select [GLP output], and then press [
OK].
(2) Select [on] or [off], and then press [
OK].
· Proceed to Step 3 if [on] has been selected.
· Proceed to Step 4 if [off] has been selected.
3.
Set the output details
(1) Select [Set], and then press [
OK].
(2) Select the items that need to be output with
[Select Details] and then press [
OK] to add the
selected item to the end of the print items listed on
the right.
(3) Repeat (2) to set all details that need to be output. To remove this setting from the print
items, press [
RIGHT], select the relevant print item and then press [
OK].
(4) When all items have been selected, press [
LEFT].
The details that can be output are listed below:
· Date
· Time
· User name (Logged in user name)*1
· User ID (Logged in user ID)*1
· Company name
· Model name
· S/N (Serial Number)
· Software version
· Balance ID
· MW (Minimum weighing value)
· Blank line
· Ruled line (----------------------)
Output settings when shipped from the
factory:
· Company name
· (Blank line)
· Model name
· S/N
· Balance ID
· (Blank line)
· Date
· Time
· (Ruled line)
*1: Output is not possible unless logged as a specific user
*2: When the minimum weighing value mode is unavailable, output is not possible even if it is set.
4.
Return to the weight measurement mode
Press [
POWER].
Precautions
Shimadzu Corp.
TYPE AP224W
SN D315300001
BALID 0000
DATE 2019 Dec.11
TIME 21.33.11
49.9921 g
-SIGNATURE-
Printer Example (with the GLP
output function set at ON)
Header
Footer