background image

Owner’s Manual

Operating Conditions

1. Mount DripAssist

2. Position DripAssist

3. Activate DripAssist

DripAssist is a monitor and alarm 

for gravity intravenous (IV) fluid 

applications. Designed for 

placement on the drip chamber of 

a tubing set, DripAssist provides 

precision feedback on IV drip 

rate. 

While you control the rate of flow 

using the roller clamp on the tubing set, DripAssist provides visual feedback, 

displaying values that accurately reflect the rate you set using the manual 

clamp. 

DripAssist is designed to be mounted directly on the drip 

chamber of a tubing set by sliding it upward. See the image 

at right.

Gently grip the drip chamber in one hand, and slide 

DripAssist upward from the tubing onto the chamber, letting 

the blue lever on the right side of the device swing 

downwards. 

The blue pads on the left and the blue lever on the right 

snugly hold the drip chamber so the DripAssist hangs 

securely.

Placing the DripAssist in the correct position on the chamber 

is made easier by the labels on the device. The orange dotted 

line indicates where you should see a free-falling drop in the 

drip chamber.

Adjust the DripAssist up or down on the drip chamber until the 

orange arrows are in line with where you see the drops fall. If 

you can see the drops in free fall at the orange arrows, then 

so can the DripAssist. 

DripAssist operates most effectively within minimum and maximum flow rates for each tubing 

set as shown in the following table: 

Drip Factor 

[gtt/mL] 

Min 

mL/hr 

Max mL/hr 

Min 

drops/min 

Max 

drops/min 

10 

12 

999 

400 

15 

999 

400 

20 

999 

400 

60 

402 

400 

If the maximum flow rate is exceeded, the device screen displays 

RATE HI.

Precautions:

When using the alarm feature, allow the flow rate to stabilize before activating the alarm 

button.

DripAssist takes one AA battery. Do not use rechargeable batteries; doing so voids the 

warranty. Ensure the battery is inserted correctly. Incorrect battery insertion can 

damage the device.

In order to ensure ongoing accuracy, DripAssist does not work when the battery level 

becomes too low. When the battery drops below a usable voltage, the screen displays 

BATTERY, and the alarm sounds. When this happens, replace the battery.

Do not 

immerse DripAssist in water. If cleaning is necessary, use a cloth and rubbing 

alcohol to wipe down the device.

Once DripAssist is correctly positioned (Steps 1 and 2), activate the device by pressing

the          (Power) button.

Any time you need to power cycle the device—such as to change a tubing set—press this button 

to turn the device off and then on again. 

4. Set The Tubing Size

As you press the        button, the screen toggles among the 
four tubing sizes, highlighting each one in turn. When you 
pause on the correct number, the DripAssist selects that 
tubing size. 

Press the        button to select the tubing size you are using. Your clinic may use any of the 
following tubing/giving sets: 

10

 gtt/mL, 

15

 gtt/mL, 

20 

gtt/mL, or 

60 

gtt/mL .

Note

: Check the packaging to see what size tubing set you’re 

using. Sometimes the tubing size is called the 

drop factor.

 

Important

: Avoid drop formation 

or splashing in the region of the 

orange indicator line.

Important

: The DripAssist 

does not 

control the rate of flow.   

Important

: This is a veterinary device only. It is not approved for use on humans.

DRIP

ASSIST

selected drop factor

Reviews: