
User Manual for Easy ECG Monitor
14
Electrode
Lead
Name
Electrode Location
Lead I
Lead II
Lead III
The intersection between
the centerline of the right
clavicle and Rib 2.
RA
RA
LA
The intersection between
the centerline of the left
clavicle and Rib 2.
LL
LA
RA
Between the left edge of the
breast bone and Rib 5
LA
LL
LL
ECG Leads Configuration and Electrodes Location
Table
(4) ECG waveform display
After the preparation, it enters the phase of ECG. The
following is the detected ECG waveform.