Microprocessor Control Water Baths 

OPERATION

 

Warning: 

These baths are not intended for use as acid baths.  Use as an acid bath will cause severe 
damage to bath components and void your warranty.  Do not use deionized water, tap water, 
or chemicals. USE DISTILLED WATER ONLY. 

Attention : 

Ces bains d'eau ne sont pas conçus pour les acides.  Utiliser un tel bain comme bain d'acide 
l'endommagera gravement et annulera la garantie.  Ne pas utiliser d'eau désionisée, d'eau 
du robinet ni de produits chimiques. UTILISER EXCLUSIVEMENT DE L'EAU DISTILLÉE. 

Turning On the Unit  

To turn on the unit, perform the following steps: 

1.  Check power supply against unit serial plate; they must match.  

2.  Plug service cord into the electrical outlet.  If supplied with a detachable cordset, plug the female end 

into the unit inlet and the male plug into the power supply.  Verify that units requiring a fuse have the 
fuse installed in the power inlet. 

3.  Fill bath to your required depth with DISTILLED WATER.  DO NOT USE TAP WATER, DEIONIZED 

WATER, OR CHEMICALS.  For singer or larger tank, normal depth is 5 ½ inches (14 cm), but depth 
must be at least 2 inches (5 cm) over the bottom of the rack.  For shallow tank, normal depth is 2/3 
full, but depth must beat least 2 inches (5 cm).  Check water level frequently, add water to appropriate 
levels if needed. At higher operating temperatures, or under circumstances where a cover cannot be 
used, it will be necessary to check the water level more frequently. 

4.  Push the Main power switch to the ON position and turn the Over Temperature Limit Thermostat to its 

maximum position, clockwise. 

Setting Main Temperature Control 

To enter set point mode on the control, push and release either the Up or Down arrow pad one time and 
the  digital  display  will  start  to  blink  from  bright  to  dim.    While  blinking,  the  digital  display  shows  the  set 
point that can be changed using the UP or DOWN arrow pads.  If the arrow pads are not pressed for five 
(5)  seconds,  the  display  will  stop  blinking  and  will  revert  to  reading  the  actual  temperature  in  the  bath.  
Allow at least two (2) hours for the temperature to stabilize. 

Warning: 

If  the  tank  boils  dry  while containing plastic ware, the plastic can melt.  If you intend to use 
test  tube  racks,  remember  that  plastic  coated  wire  racks  may  wear  and  expose  metal  that 
can cause damage.  Preferably, use all plastic racks. 

Attention : 

Si la cuve vient à s'assécher tandis qu' elle contient des ustensiles en plastique, le plastique 
risque  de  fondre.    Si  des  râteliers  à  tubes  à  essais  constitués  d'une  ossature  métallique 
gainée  de  plastique  sont  utilisés,  prendre  garde  que  le  plastique  peut  fondre  ou  s'user  et 
découvrir  des  parties  métalliques  risquant  d'endommager  différentes  pièces.    Il  est 
préférable d'utiliser des râteliers à tubes entièrement en plastique. 

Calibrating the Main Temperature Control 

We recommend that you calibrate your unit once it has been installed in its working environment and the 
chamber temperature has been stable at the set point for several hours.  

Section 

Summary of Contents for SWB15

Page 1: ...2 SWB15 SWB15 2 SWB23 SWB23 2 SWB715 SWB 715 2 Previously known as W2M W2M 2 W6M W6M 2 W14M W14M 2 W20M W20M 2 W614M W614M 2 115v 220v and PC Models MICROPROCESSOR CONTROLLED INSTALLATION AND OPERATIO...

Page 2: ...ile or combustible materials are being heated These units have been tested to the following requirements CAN CSA C22 2 No 61010 1 2012 CAN CSA C22 2 No 61010 2 010 R 2009 UL 61010 1 2004 R 2005 07 R 2...

Page 3: ...amination 7 CONTROLS OVERVIEW 8 Power Switch 8 Main Temperature Controller 8 Fuses 8 Over Temperature Limit Thermostat OTL 8 Heating Activated 8 Over Temperature Activated Light 8 OPERATION 9 Turning...

Page 4: ...nual before using the water bath Be sure you follow any city county or other ordinances in your area regarding the use of this unit Use only approved accessories Do not modify system components Any al...

Page 5: ...delivered the unit If the carton is not damaged open the carton and remove its contents Verify that all of the following equipment is included in the crate One 1 water bath One 1 bath cover two 2 for...

Page 6: ...p re secteur AC I Indicates the power is ON Rep re de la position MARCHE de l interrupteur d alimentation O Indicates the power is OFF Rep re de la position ARR T de l interrupteur d alimentation Indi...

Page 7: ...ts are intended for 50 60 HZ application A separate circuit is recommended to prevent damage to the unit due to overloading or circuit failure Location In selecting a location consider all conditions...

Page 8: ...uit breaker and will cut off power to the unit if there is an electrical surge or malfunction The cause should be determined and corrected before placing the fuse Fuses must be in place for the unit t...

Page 9: ...ll be necessary to check the water level more frequently 4 Push the Main power switch to the ON position and turn the Over Temperature Limit Thermostat to its maximum position clockwise Setting Main T...

Page 10: ...Thermostat To set the Over Temperature Limit Thermostat perform the following steps 1 Verify that the Limit Thermostat was set to its maximum position to allow the water bath to stabilize 2 Turn the...

Page 11: ...ak through openings and cracks and get on electrical components or that may contain solvents corrosives or abrasives that will harm the stainless steel coatings 2 If a hazardous material substance has...

Page 12: ...g temperatures or in applications where a cover is not used more frequent water level checking is required Test tube Racks If the water bath tank boils dry while containing plastic ware the plastic wi...

Page 13: ...n Temperature too high Insufficient quantity of water Fill bath a minimum of three fourths 3 4 full with water Main controller set too high See Setting Main Temperature Control Main controller failed...

Page 14: ...aintain set point Insufficient quantity of water Fill bath a minimum of three fourths 3 4 full with water Temperature set too low Assure that set point is at least 5 degrees over ambient room temperat...

Page 15: ...B7 2 W6M 2 9751191 9751191 Gable Covers SWB15 W14M SWB15 2 W14M 2 9751184 9751184 Gable Covers SWB23 W20M SWB23 2 W20M 2 9751185 9751185 Cable Covers SWB715 W614M SWB715 2 W614M 2 9751191 9751184 9751...

Page 16: ...13 Tank Gasket 3450538 3450538 Tanks SWB2 W2M SWB2 2 W2M 2 7930521 7930521 Tanks SWB7 W6M SWB7 2 W6M 2 SWB715 W614M SWB715 2 W614M 2 7930523 7930523 Tanks SWB15 W14M SWB15 2 W14M 2 SWB715 W614M SWB715...

Page 17: ...1 in 37 47 x 31 12 x 30 73 cm 11 8 x 5 8 x 2 5 in 29 98 x 14 73 x 6 35 cm SWB7 SWB72 2 W6M W6M 2 14 75 x 12 25 x 12 1 in 37 47 x 31 12 x 30 73 cm 11 8 x 5 8 x 6 in 29 98 x 14 73 x 15 24 cm SWB15 SWB1...

Page 18: ...02 W20M W20M 2 14 17 SWB715 SWB715 2 W614M W614M 2 4 7 small tank 11 225 large tank 5 75 small tank 14 75 large tank Model Range Uniformity Sensitivity SWB2 SWB2 2 W2M W2M 2 Amb 5 C to 80 C 2 C 37 C 1...

Page 19: ...240 V 2 A 50 60 Hz SWB7 SWB72 2 W6M W6M 2 Volts 100 120 V 3 A 50 60 Hz Volts 220 240 V 2 A 50 60 Hz SWB15 SWB15 2 W14M W14M 2 Volts 100 120 V 5 A 50 60 Hz Volts 220 240 V 3 A 50 60 Hz SWB23 SWB2302 W2...

Page 20: ...505 POWER LIGHT 4650554 TEMPERATURE CONTROL 1750960 OTP LIGHT 4650553 HEATING LIGHT 4650554 OTP THERMOSTAT 1750747 300W 41 8 BLACK CHERRY ROUND POWER SWITCH 7850553 HOT TERMINAL BLOCK NEUTRAL TERMINAL...

Page 21: ...4 TEMPERATURE CONTROL 1750961 OTP LIGHT 4650553 HEATING LIGHT 4650554 OTP THERMOSTAT 1750747 300W 153 6 BLACK CHERRY ROUND POWER SWITCH 7850553 HOT TERMINAL BLOCK NEUTRAL TERMINAL BLOCK 2800502 EMI FI...

Page 22: ...302 HOT NEUTRAL GROUND LOAD SENSOR 1 2 4 FUSED INLET 4200505 POWER LIGHT 4650554 TEMPERATURE CONTROL 1750957 OTP LIGHT 4650553 HEATING LIGHT 4650554 OTP THERMOSTAT 1750747 300W 41 8 BLACK CHERRY ROUND...

Page 23: ...NEUTRAL GROUND LOAD SENSOR 1 2 4 FUSED INLET 4200505 POWER LIGHT 4650554 TEMPERATURE CONTROL 1750958 OTP LIGHT 4650553 HEATING LIGHT 4650554 OTP THERMOSTAT 1750747 300W 153 6 BLACK CHERRY ROUND POWER...

Page 24: ...1304 HOT NEUTRAL GROUND LOAD SENSOR 1 2 4 FUSED INLET 4200505 POWER LIGHT 4650554 TEMPERATURE CONTROL 1750955 OTP LIGHT 4650553 HEATING LIGHT 4650554 OTP THERMOSTAT 1750747 500W 24 5 BLACK CHERRY ROUN...

Page 25: ...T NEUTRAL GROUND LOAD SENSOR 1 2 4 FUSED INLET 4200505 POWER LIGHT 4650554 TEMPERATURE CONTROL 1750956 OTP LIGHT 4650553 HEATING LIGHT 4650554 OTP THERMOSTAT 1750747 600W 90 1 BLACK CHERRY ROUND POWER...

Page 26: ...1306 HOT NEUTRAL GROUND LOAD SENSOR 1 2 4 FUSED INLET 4200505 POWER LIGHT 4650554 TEMPERATURE CONTROL 1750953 OTP LIGHT 4650553 HEATING LIGHT 4650554 OTP THERMOSTAT 1750747 500W 24 5 BLACK CHERRY ROUN...

Page 27: ...T NEUTRAL GROUND LOAD SENSOR 1 2 4 FUSED INLET 4200505 POWER LIGHT 4650554 TEMPERATURE CONTROL 1750954 OTP LIGHT 4650553 HEATING LIGHT 4650554 OTP THERMOSTAT 1750747 600W 90 1 BLACK CHERRY ROUND POWER...

Page 28: ...LIGHT 4650554 OTP THERMOSTAT 1750747 500W 24 5 BLACK CHERRY ROUND POWER SWITCH 7850553 HOT TERMINAL BLOCK NEUTRAL TERMINAL BLOCK HOT NEUTRAL GROUND LOAD SENSOR 1 2 4 POWER LIGHT 4650554 TEMPERATURE C...

Page 29: ...4650554 OTP THERMOSTAT 1750747 600W 90 1 BLACK CHERRY ROUND POWER SWITCH 7850553 HOT TERMINAL BLOCK NEUTRAL TERMINAL BLOCK 2800502 EMI FILTER HOT NEUTRAL GROUND LOAD SENSOR 1 2 4 POWER LIGHT 4650554 T...

Reviews: