11
MAINTENANCE
Warning
: Prior to any maintenance or service on this unit, disconnect the power cord from the power source.
Before reattaching the unit to its power source, be sure all volatile and flammable cleaners are
evaporated and dry.
Avertissement
: Avant d'effectuer toute maintenance ou entretien de cet appareil, débrancher le cordon
secteur de la source d'alimentation. Avant de reconnecter l'appareil sur le secteur, s'assurer
que tous les produits de nettoyage volatiles et inflammables sont complètement évaporés.
The design of the chamber is such that periodic maintenance is kept to a minimum. No lubrication or adjustments of
components is needed.
Note:
The unit chamber should be cleaned and disinfected prior to use.
Cleaning:
Periodic cleaning is required. To clean the incubator, perform the following steps:
1.
Remove all of the interior parts, if assembled.
2.
Clean the incubator with a mild soap and water solution, including all corners. DO NOT USE spray
cleaners that might leak through openings and cracks and get on electrical components, or that may
contain solvents that will harm coatings. DO NOT USE chlorine-based bleaches or abrasives, as they
will damage the painted interior.
3.
Rinse with distilled water and wipe dry with a soft cloth.
Special care should be taken when cleaning around the sensing heads to prevent damage.
7.2
Disinfecting:
Disinfect the incubator on a regular basis. To disinfect the incubator, perform the following steps.
1. Remove all of the interior parts, if assembled.
2. Disinfect the incubator, including all corners, using a suitable disinfectant. Shelves and shelf clips are
autoclaveable. DO NOT USE spray disinfectants that might leak through openings and cracks and get
on electrical components, or that may contain solvents, corrosives, or abrasives that will harm the
painted coatings. Special care should be taken when cleaning around sensing heads to prevent
damage and around the door gasket so as not to impair the positive seal.
3.
If a hazardous material/substance has been damaged in the unit, immediately initiate your site’s
Hazardous Material Spill Containment protocol. Contact your local Site Safety Officer and follow
instructions per the policy and procedures established for your site.
4. There are many commercially available disinfectants available that are non-corrosive and non-abrasive
and suitable for use on painted surfaces. Contact your local Site Safety Officer for detailed information
for the proper disinfectants suitable for your operation.
Warning
:
Never clean the unit with alcohol or flammable cleaners and assure all volatile or flammable cleaners
are evaporated and dry before reattaching the unit to the power supply.
Avertissement
:
Ne jamais nettoyer l'appareil à l'alcool ou avec des nettoyants inflammables et veiller à ce que
les produits volatils ou inflammables soient entièrement évaporés avant de rebrancher le
content d'alimentation de l'appareil.
Periodically inspect the door latch, trim, catch and gasket for signs of deterioration. Failure to maintain the integrity of
the door system will shorten the life span of the incubator.
No maintenance is required on electrical components. If the incubator fails to operate as specified, please review
the TROUBLESHOOTING section prior to calling for service.
Section
7