18
|
P a g e
INSTALLATION
D
EIONIZED AND
D
ISTILLED
W
ATER
Do not use deionized water
to clean the chamber even if it is readily available in your laboratory
environment. Use of deionized water may corrode metal surfaces and voids the warranty. Sheldon
Manufacturing recommends the use of distilled water in the resistance range of 50K Ohm/cm to 1M
Ohm/cm, or a conductivity range of 20.0 uS/cm to 1.0 uS/cm, for cleaning applications.
I
NSTALLATION
C
LEANING AND
D
ISINFECTION
The manufacturer recommends cleaning the shelving and unit chamber prior to installation of the
shelving in the chamber. The unit was cleaned at the factory but may have been exposed to
contaminants during shipping. Remove all wrappings and coverings from shelving prior to cleaning
and installation.
Do not clean with deionized water.
Cleaning and Disinfection procedure
on page 29 of the User Maintenance section
for information on how to clean and disinfect without damaging the unit and its components.