38
|
P a g e
OPERATION
M
UTING THE
A
UDIBLE
CO
2
A
LARM
A visual devation indicator alarm illuminates if the chamber CO
2
level deviates 1% above or below the
CO
2
set point. An audible alarm sounds immediately for a high deviation. The low deviation audible
alarm sounds only after the visual low indicator light has been continually illuminated for 15 minutes.
This delay prevents the alarm from sounding each time the chamber doors are opened.
To mute a CO
2
deviation alarm:
Press and hold the Set CO
2
Up
or
Down
arrow button until the Mute light
illuminates.
•
The alarm will stay muted for the duration of the current
temperature deviation.
•
A new deviation of 1% or more will reactivate the audible alarm.
Low and No Gas Alarms Timeline
N
O
G
AS
S
UPPLY
A
LARM
A second alarm activates if the Low Gas deviation indicator light is illuminated for longer
than 20 minutes.
•
The letters “ngS” will appear in the CO
2
display indicating
No Gas Supply
.
•
The alarm will remain active, even if the incubator is turned off and back on.
•
The NGS alarm remains on until CO
2
is restored to the chamber. It may take 2 or more
minutes of CO
2
injections to establish a concentration high enough to deactivate the
alarm.
Low Gas Alarm Muted
No Gas Supply
Summary of Contents for SCO5W
Page 1: ...WATER JACKETED CO2 INCUBATOR 110 120 Volts Installation Operation Manual SCO5W ...
Page 2: ...2 P a g e ...
Page 6: ...6 P a g e TABLE OF CONTENTS ...
Page 14: ...14 P a g e RECEIVING YOUR UNIT ...
Page 24: ...24 P a g e INSTALLATION ...
Page 56: ...56 P a g e USER MAINTENANCE ...
Page 61: ...61 P a g e PARTS LIST ...