INTRODUCTION
R
EFERENCE
S
ENSOR
D
EVICES
Must be purchased separately
Reference sensor devices or a combined device are required for calibrating the
incubator temperature and CO
2
displays.
Reference devices must meet the following standards:
•
Accurate to at least 0.1°C
•
Accurate to at least 0.1% CO
2
The devices should be regularly calibrated, preferably by a third party.
Temperature Probes
For temperature, use a digital device with a wire thermocouple probe that can be
introduced into the incubation chamber through the unit access port. Select a
thermocouple suitable for the application temperature you will be calibrating at.
CO
2
Sampling
For best CO
2
accuracy, use a digital gas analyzer with sample tubing connecting to
the incubator CO
2
sample port. The barbed adapter gas fitting on the port connects
to
3/16-inch (4.76mm)
inner diameter (ID),
5/16-inch (7.94mm)
outside diameter (OD)
tubing.
Why Probes and Tubing?
Reference readings taken outside the chamber using wire temperature probes and
CO
2
drawn through the sample port avoid chamber door openings. Openings disrupt
the chamber temperature and CO
2
concentration level. Each disruption requires
a
minimum 1-hour wait
to allow the atmosphere to re-stabilize before continuing.
No Alcohol or Mercury Thermometers
Alcohol thermometers do not have sufficient accuracy to conduct accurate
temperature calibrations.
Never place a mercury thermometer in the incubation
chamber.
Always use thermocouple probes.
CO
2
Reference
Temperature
Reference
Summary of Contents for SCO5W
Page 1: ...WATER JACKETED CO2 INCUBATOR 110 120 Volts Installation Operation Manual SCO5W ...
Page 2: ...2 P a g e ...
Page 6: ...6 P a g e TABLE OF CONTENTS ...
Page 14: ...14 P a g e RECEIVING YOUR UNIT ...
Page 24: ...24 P a g e INSTALLATION ...
Page 56: ...56 P a g e USER MAINTENANCE ...
Page 61: ...61 P a g e PARTS LIST ...