19 |
P a g e
INSTALLATION
I
NSTALL THE
I
NCUBATOR
Install the unit in a workspace location that meets the criteria discussed in the previous entries of the
Installation section.
D
EIONIZED AND
D
ISTILLED
W
ATER
Do not use deionized water
to clean the unit, even if DI water is readily available in your laboratory.
•
The use of deionized water may corrode metal surfaces and
is not covered by the
manufacturing defect warranty
.
•
The manufacturer recommends the use of distilled water in the resistance range of 50K
Ohm/cm to 1M Ohm/cm, or a conductivity range of 20.0 uS/cm to 1.0 uS/cm, for cleaning
applications.
I
NSTALLATION
-
C
LEAN AND
D
ISINFECT
The manufacturer recommends cleaning the shelving, ceiling air duct, and chamber prior to Installing
the shelving and ceiling duct in the chamber.
•
The unit was cleaned at the factory but may have been exposed to contaminants during
shipping.
•
Remove all wrappings and coverings from shelving and the ceiling air duct prior to cleaning
and installation.
Do not clean the shelving with deionized water.
•
Do not clean the chamber HEPA filter.
•
Please see the
procedure on page 43 in the User Maintenance
chapter for information on how to clean and disinfect without damaging the unit.