25
R-209HK
R-209HW
DOOR REPLACEMENT
1. Disconnect the power supply cord, and then remove
outer case.
2. Open the door and block it open.
3. Discharge high voltage capacitor.
If the door sensing switch, secondary interlock switch and
monitor switch do not operate properly due to a
misadjustment, the following adjustment should be made.
4. Loosen the two (2) screws holding latch hook to the oven
cavity front flange.
5. With door closed, adjust latch hook by moving it back and
forth, and up and down. In and out play of the door
allowed by the upper and lower position of the latch hook
should be less than 0.5mm. The vertical position of the
latch hook should be adjusted so that the door sensing
switch and secondary interlock switch are activated with
the door closed. The horizontal position of the latch hook
should be adjusted so that the plunger of the monitor
switch is pressed with the door closed.
6. Secure the screws with washers firmly.
7. Check the operation of all switches. If each switch has
not activated with the door closed, loosen screw and
adjust the latch hook position.
After adjustment, check the following.
1. In and out play of door remains less than 0.5mm when in
the latched position. First check upper position of latch
DOOR SENSING SWITCH/SECONDARY INTERLOCK SWITCH AND MONITOR SWITCH ADJUSTMENT
hook, pushing and pulling upper portion of door toward
the oven face. Then check lower portion of the latch
hook, pushing and pulling lower portion of the door
toward the oven face. Both results (play in the door)
should be less than 0.5mm.
2. The door sensing switch and secondary interlock switch
interrupt the circuit before the door can be opened.
3. Monitor switch contacts close when door is opened.
4. Re-install outer case and check for microwave leakage
around door with an approved microwave survey meter.
(Refer to Microwave Measurement Procedure.)
Figure C-4. Latch Switch Adjustments
Latch Heads
Latch Hook
Door
Door
Sensing
Switch
Monitor
Switch
Switch Lever
Secondary
Interlock
Switch
REMOVAL
1. Disconnect the power supply cord.
2. Open the door slightly.
3. Insert a putty knife (thickness of about 0.5mm) into the
gap between the choke cover and corner portion of door
panel as shown in Figure C-5 to free engaging parts.
4. Pry the choke cover by inserting a putty knife in order
shown in figure C-5.
5. Release choke cover from door panel.
6. Now choke cover is free.
NOTE: When carrying out any repair to the door, do not
bend or warp the slit choke (tabs on the door
panel assembly) to prevent microwave leakage.
Figure C-5. Door Disassembly
7. Release two (2) pins of door panel from two (2) holes of
upper and lower oven hinges by lifting up.
8. Now, door panel is free from oven cavity.
9. Release door panel from ten (10) tabs of door frame and
remove door frame by sliding the door panel downward.
10.Now, door panel with sealer film is free.
11.Tear sealer film from door panel.
12.Now, door panel is free.
13.Slide latch head upward and remove it from door frame
with releasing latch spring from door frame and latch
head.
14.Now, latch head and latch spring are free.
15.Remove door screen from door frame.
16.Now, door screen is free.
RE-INSTALL
1. Re-install door screen to door frame.
2. Re-install latch spring to the head. Re-install latch spring
to the door frame. Re-install latch head to the door frame.
3. Re-install door panel to door frame by fitting ten (10) tabs
of door frame to ten (10) holes of door panel.
4. Put sealer film on door panel. Refer to "Sealer Film" and
figure C-7, on how to handle the new film.
10
5
4
8
7
9
11
3
2
1
Choke Cover
Door Frame
Putty Knife
6
Re-install
1. Re-install each switch in its place. The secondary
interlock/monitor switches are in the lower position and
the door sensing switch is in the upper position.
2. Re-connect wire leads to each switch.
Refer to pictorial diagram.
3. Secure latch hook (with two (2) mounting screws) to
oven flange.
4. Install the switch lever to the oven cavity
5. Make sure that the monitor switch is operating properly
and check continuity of the monitor circuit. Refer to
chapter "Test Procedure" and Adjustment procedure.