SGS Household Oxygen Bar Operating Instructions Manual Download Page 8

10

Problem Investigation and Trouble Shooting

Symptom

Possible Cause

Solution

Household oxygen
bar cannot operate
properly, the power
indicator light is not
bright.

The power line is not conne-
cted to a power outlet.

Power supply socket without 

electricity.

Overload protection burning fuse

Power cord is not for good in
 the household oxygen bar or
 power cord bad.

Check whether the socket outlet connection 
is normal.

Replace the electric socket.

Open the insurance and replace the fuse.

Unplug reinsert the fuselage outlet or repl-
ace a power supply cord.

The screen display 
normal, but the oxy-
gen bar does not ha-
ve normal operation 
soun.

Household oxygen bar over-
heating protection.

Household oxygen bar com-
puter froze plate.

The compressor function b-
reakdown.

Make sure the place has been away from ho-
usehold oxygen bar curtains, bake warm m-
achines, heaters and fireplaces, and confirm 
household oxygen bar and walls or other o-
bjects around the retention gap of at least 
20 cm. 

Turn off the machine, wait five minutes to
 restart. 

Contact household oxygen bar retailer or m-
anufacturers.

Full screen flash, 
while alarm. 

Internal hardware damage 

Returned to the manufacturer or dealer for 
repair.

Big Noise 

Ties are not cut, the machine
did not put a smooth. 

Cut ties, put the machine steady.

The screen display 
normal and hear n-
ormal operation so-
und but no anoin o-
xygen output. 

Intake window cotton filters
is dirty, cause block.

Household oxygen bar internal malfunction. 

Cut off the power switch, r-
eplace window air intake fi-
lter cotton. 

Contact the household seller or manufactu-
rers.

Press to adjust the
time when the tra-
nsfer is invalid. 

Sleep tone when keys are lo-
cked, inoperable. 

Press the "On / Off" button to wake, the key
to adjust the time and then tune.

09

Problem Investigation and Trouble Shooting

Symptom

Possible Cause

Solution

The oxygen machine
 emits poof bared  s-
ound six-minute int-
ervals.

Water coming out 
of the mouth of o-
xygen.

the machine is op-
erating normally -
after starting up, 
but have a narrow
minded.

Oxygen odor

The machine emits exhaust 
sound.

Adding too much water in the
 wet cup

exceed the highest

 water level

),

airflow will 

push oxygen tube blisters.

Difference temperature , ca-
use the humidified oxygen in
the wall that moisture will c-
ondense water out of cold. 

Flow value is too small.

Oxygen tube was twisted or
 flatten.

Early use, series of tubes or 
other plastic will emit odor,
 is a normal phenomenon.

If the machine emits poof bared sound reg-
ularly is a normal phenomenon, discharge
the sound of nitrogen.

The water in the wet cup have reduced to the
 highest water level.

Pour the water out of the wet cup, starting 
up, after fifteen minutes ,waiting the dropl-
ets are dried in the oxygen tube, then adding
 water into the wet cup.

Turn up flow

1. Tank silicone stopper loose, remove the 
tank and the silica gel plug impaction.

2. The tank did not put into the machine in
place, should be put into the tank with a lit-
tle push.

3. The gas tank has two aprons in and out of
the mouth, if not installed well, it can lead 
to leakage.

This odor is normal smell emitted by the sil-
icone tube, which has been detected by non
toxic, and with the use of the machine, the
 smell will gradually disappear.

Summary of Contents for Household Oxygen Bar

Page 1: ...ygen flow Cumulat ivetime Spray Power supply Anion O2 Oxygenco ncentration Oxygen Minute Oxygen flow Cumulat ivetime Spray Power supply Anion Operating Instructions Household Oxygen Bar The specificat...

Page 2: ...warranty Be careful the company reserves the rights to update pro ducts and technology Some illustrations in the specifica tion you see in the product may be different received in kind prevail Thank y...

Page 3: ...ng household oxygen bar by the cord or hose is prohibited This product with rated voltage of 220V please use secure socket qualified To prevent accidents do not let children operate the oxygen bar Whe...

Page 4: ...e the user operate it comfortably Temperature 5 35 Relative humidity 80 Range of the atmospheric pressure 0 086 MPa 0 106 MPa When the temperature is higher or lower do not use oxygen bar 5000 hours l...

Page 5: ...er Power switch Insurance holder Power cord socket Cooling air outlet Cooling air inlet O2 Oxygen co ncentration Oxygen Minute Oxygen flow Cumulat ive time Power supply Anion Installation and Operatio...

Page 6: ...on see section appearance of page four Silicon capsule Highest lines Installation and Operation O2 Power on ON OFF button In standby mode press the ON OFF to wake up the machine and start wo rking and...

Page 7: ...upply and unplug the power in time The cotton filters on the host side change in time if the cotton filters we re blocked by the dirty stuff it will have an effect on oxygen Attention Please don t ta...

Page 8: ...window air intake fi lter cotton Contact the household seller or manufactu rers Press to adjust the time when the tra nsfer is invalid Sleep tone when keys are lo cked inoperable Press the On Off but...

Page 9: ...ler Coarse filter level 1 Dust filter level 2 Oil free air compressor Nitrogen exclude silencer One piece molecules containing separation valve tower Bacterial filter Wet cup Oxygen outlet Electronic...

Page 10: ...13 ISO13485...