Page 22 of 32
www.sgm-lektra.com
8.5
“DIAGNOSTIC” menu
8.6 - DIAGNOSTIC
From “RUN” mode, holding down UP ARROW, press ENTER to access
Position the cursor on DIAGNOSTIC and press ENTER
Select the parameters by moving the cursor with SCROLL and con
fi
rm with ENTER
VLW602 -
advanced con
fi
guration
8.6.1 - ALARM CONFIGURATION
Position the cursor on ALARM CONFIGURATION, press ENTER to access
To enable or disable each diagnostic alarms.
- with UP ARROW or SCROLL chose the desired item and press ENTER
- with UP ARROW or SCROLL enable or disable the alarm signal and
press ENTER to con
fi
rm
DIAGNOSTIC
ALARM CONFIGURATION
MEASURE STATUS
FROZEN GAIN
MAX GAIN TH.
PEAK VALUES
OUTPUT SIMUL
TEMPERATURE
NO ECHO FOUND
MAX GAIN
ECHO IN BLIND
DISTANCE >120%
MEASURE STATUS
G : 00000
FROZEN GAIN
000
MAX GAIN TH
200
>DISPLAY VALUES
RESET VALUES
OUTPUT SIMUL
00.0mA
DISABLE
>ENABLE
PEAK VALUES
Max 0000mm
Min 0000mm
ALARM CONFIGURATION
MEASURE STATUS
FROZEN GAIN
MAX GAIN TH.
PEAK VALUES
OUTPUT SIMUL.
Ź
SETUP
DISPLAY
DIAGNOSTIC
SERVICE
PROBE INFO
UNIT INFO
Ź
ALARM CONFIGURATION
MEASURE STATUS
FROZEN GAIN
MAX GAIN TH.
PEAK VALUES
OUTPUT SIMUL.
Ź
TEMPERATURE
NO ECHO FOUND
MAX GAIN
ECHO IN BLIND
DISTANCE >120%
Ź
DISABLE
ENABLE
Ź