49
6 Faults
6
Faults
If a fault occurs during operation, an error message will appear
on the screen.
Error messages are displayed in the order in which they occur.
Up to five errors can be displayed. Error messages continue to
be displayed until the error is corrected.
Overview of possible error messages
Error
code
Error message on the
display
Error correction
8
No calibration
PPM sensor adjustment
H2 ppm adjustment required
)
10
Adjustment failed
Test gas
Check test gas
)
52
XFLASH
SEWERIN Service
Error can only be corrected by
SEWERIN Service
59
Error unknown
SEWERIN Service
Error can only be corrected by
SEWERIN Service
60
PX sensor
Error can only be corrected by
SEWERIN Service
100
Pump error
Probe/filter
Check all filters, probes and hose
connections for porosity and dirt
202
I2C HOST – EX
SEWERIN Service
Error can only be corrected by
SEWERIN Service