6
│ 2 Features
NOTICE! Damage caused by leaking sensors
Electrochemical sensors that are faulty or whose expected ser-
vice life is exceeded can leak. The leaking substance can cause
permanent damage to the device.
Faulty sensors can be identified, for example, by the fact that the
corresponding adjustment is no longer possible.
The installation date of the sensor and the expected lifetime can
be found in the
Device information
.
●
Always arrange for SEWERIN Service to replace faulty or leaky
sensors immediately.
2.3
Explosion protection
The device is assigned to the following explosion-proof groups:
Explosion-proof
group
For the following
atmospheres
When using
II2G Ex d e ib IIB T4 Gb
Methane CH
4
Propane C
3
H
8
Butane C
4
H
10
Hydrogen sulphide H
2
S
Device
without
carrying bag
TG8
II2G Ex d e ib IIC T4 Gb
Methane CH
4
Propane C
3
H
8
Butane C
4
H
10
Hydrogen sulphide H
2
S
Hydrogen H
2
Device
with
carrying bag
TG8
EC type-examination certificate: TÜV 07 ATEX 553353 X
A
DANGER! Risk of explosion from sparks
●
Only ever open the battery compartment outside of ex-
plosive areas.
●
Only ever charge the device outside of explosive areas.
●
Only use the USB port outside of explosive areas.
●
Use only suitable battery types.