Setra Systems 205-2 Operating Instructions Download Page 2

RETURNING  PRODUCTS FOR REPAIR

Please contact Setra (1-800-257-3872 or 978-263-1400) before returning unit for repair to review information 

relative to your application.  Many times, only minor field adjustments may be necessary.

When returning a product to Setra, the material should be carefully packaged and shipped prepaid to:
    Setra 

Systems, 

Inc.

    159 

Swanson 

Road

    Boxborough, 

MA 

01719

    Attn: 

Repair 

Department

To assure prompt handling, please supply the following information and include it inside the package of 

returned material:
 

 

 

 

1.  Name and phone number of person to contact.

 

 

 

 

2.  Shipping and billing instructions.

 

 

 

 

3.  Full description of the malfunction.

 

 

 

 

4.  Identify any hazardous material used with product.

Notes: Please remove any pressure fittings and plumbing that you have installed and enclose any required 

mating  electrical connectors and wiring diagrams.
Allow approximately 3 weeks after receipt at Setra for the repair and return of the unit.

Non-warranty repairs will not be made without customer approval and a purchase order to cover repair 

charges.

in to the output circuit to cause output oscillation.  If encountered, this oscillation may be eliminated by 

connecting a 100 ohm resistor (1/8 watt or larger) in series in each of the output leads at the end of the 2 

foot transducer cable.  These series resistors, of course, add to the output resistance.

CAUTION:  

Excitation power, or voltage in excess of 15 VDC, inadvertently applied to the output leads 

may damage the electrical circuit.  Care must be taken when installing this transducer that the excitation 

voltage is not applied to the output leads by mistake.  Shielding or other precaution should be provided 

to assure that transient voltages in excess of 15 VDC are not applied to the output leads.

ADJUSTMENTS (with cover removed)
CAUTION:  TURN OFF EXCITATION POWER DURING COVER REMOVAL OR REPLACEMENT.
ZERO PRESSURE OUTPUT

Unit factory adjusted to within ± 50 mV.  Can be adjusted by the air trim capacitor as shown in the dia-

gram.  Any adjustments will begin to change sensitivity.

SENSITIVITY

Can be adjusted by potentiometer as shown on diagram.  Unit factory adjusted to order specifications.

OTHER ADJUSTMENT

Adjusted at factory.  Touching any adjustments other than zero output or sensitivity may necessitate 

recalibration.  Do not touch.

INSTALLATION INSTRUCTIONS FOR MULTIPLE HOOK UP OF A 4-WIRE SETRA PRESSURE TRANSDUCER

CALIBRATION SERVICES

Setra maintains a complete calibration facility that is traceable to the  National Institute of Standards & Technol-

ogy (NIST).  If you would like to recalibrate or recertify your Setra pressure transducers please call our Repair 

Department at 1-800-257-3872 (978-263-1400) for scheduling, cost and turnaround estimates.

Option #1

Uses a single, ground referenced power supply 

for excitation and separate, isolated, (not ground 

referenced), readout or "differential input" to a 

data acquisition system for each pressure trans-

ducer output.

Option #3

Uses a single, ground referenced power supply

for excitation and either a single, isolated readout 

with a bipolar switch that "breaks before makes" 

both the + output and - output of each pressure 

transducer, or a single data acquisition system 

with a multiplexer  (MUX).

NOTE:  The shield is internally commoned to the 

case and pressure port of the transducer.  When 

the shield is connected to ground, the case and 

pressure port of the transducer will also be com-

moned to that ground.

PS - Power Supply (Nominal 24 VDC)

PT - Pressure Transducer (4 Wire Circuit)

DVM/DAS - Digital Volt Meter or Data Acquisition 

System

Option#2

Uses a separate, isolated power supply for each 

pressure transducer's excitation and a single, 

ground referenced readout or "single-ended 

input" to a data acquisition system for all of the 

outputs.

LIMITED WARRANTY AND LIMITATION OF LIABILITY

SETRA warrants its products to be free from defects in materials and workmanship, subject to the following terms and 

conditions:  Without charge, SETRA will repair or replace products found to be defective in materials or workmanship 

within the warranty period; provided that:

a)   the product has not been subjected to abuse, neglect, accident, incorrect wiring not our own, improper instal-

lation or servicing, or use in violation of instructions furnished by SETRA;

b)   the product has not been repaired or altered by anyone except SETRA or its authorized service agencies;
c) 

the serial number or date code has not been removed, defaced, or otherwise changed; and

d)   examination discloses, in the judgment of SETRA, the defect in materials or workmanship developed under 

normal installation, use and service;

e)   SETRA is notified in advance of and the product is returned to SETRA transportation prepaid.

Unless otherwise specified in a manual or warranty card, or agreed to in writing and signed by a SETRA officer, SETRA 

pressure and acceleration products shall be warranted for one year from date of sale.
The foregoing warranty is in lieu of all warranties, express, implied or statutory, including but not limited to, any implied 

warranty of merchantability for a particular purpose.
SETRA’s liability for breach of warranty is limited to repair or replacement, or if the goods cannot be repaired or replaced, 

to a refund of the purchase price.  SETRA’s liability for all other breaches is limited to a refund of the purchase price.  In 

no instance shall SETRA be liable for incidental or consequential damages arising from a breach of warranty, or from the 

use or installation of its products.
No representative or person is authorized to give any warranty other than as set out above or to assume for SETRA any 

other liability in connection with the sale of its products.