28
13
Remove the lid from the photometer.
14
Open the <PUMPS> tab.
15
Press the <ON> button next to “P2 - Salicylate”.
16
While the pump is running, look into the photometer and wait until
the reagent arrives in the photometer.
This takes approximately 2.5 minutes. The pump stops
automatically.
17
Go back to the <ACTUATOR> tab.
18
Set “Stirrer” to <ON>.
19
Check that the stirrer bar is turning.
20
Reinstall the lid on the photometer.
ON
ON
ON
ON
OFF
OFF
OFF
OFF
P1 - Buffer
P1 - Buffer
P2 - Salicylate
P3 - Dichloro Acid
P9 - H2SO4 2N
ACTUATOR
4-20mA OUTPUT
PUMPS
RELAYS
INPUTS