combination with the conductor-plating (gold, nickel, etc.) and roughness
may lead to a slightly (
≈
0.2) increased dielectric constant.
Figure 12: Result visualizations for
ε
plus
mode
In the right image in Fig.
the frequency dependent dielectric ”constant”
ε
r
(
f
)
is displayed versus frequency. The left image shows line impedance
Z
0
(
f
)
versus the frequency.
Please note:
it can be advantageous to enable frequency domain smoothing (see
chapter
6.7 Permittivity Measurements (optional, not for DTDR-800)
The permittivity measurement mode is designed to work with an open-en-
ded coaxial line sensor (e.g. SDM-10G of Sequid GmbH ). This mode was
developed for material measurement applications and it is essential to acti-
vate the
TTC
(see chapter
•
Right-click an existing sample and select the permittivity mode. In
case it is not available, please
Sequid GmbH for further infor-
mation.