the zero point (
x
= 0
cm) to a well defined value.
Please note:
the propagation velocity according to the DUT must be set in
the Impedance Measurement Properties Window (see section
•
An automatic zoom can be activated by checking the
Auto Zoom
check-
box or via the menu
View
→
Auto Zoom
.
•
In time domain measurement modes (
TDR, DTDR
) the zoom can be
locked to the
Oscilloscope Window
zoom by pressing the toolbar but-
ton with the lock symbol.
•
The impedance tolerance mask can be set via the
Extra
→
Set…
menu
items. For further information see section
about the
Impedance
Measurement Properties Window
.
•
Optional:
For an increased accuracy the TTC should be applied. If
TTC measurements have been performed, the
corrected impedance-
or TDR-traces
can be displayed by checking the
Correction
checkbox
in the
Result Visualization Window
. For a detailed description about
working with TTC, please refer to chapter
.
•
Please note:
other rise times than the predefined ones can be simu-
lated if TTC is activated. This option is capable of characterizing sys-
tems with lower and higher bandwidths. For simulating higher band-
widths (lower rise time) an appropriate number of averages is recom-
mended.
•
Please note:
Impedance calculation and TTC are not applicable to
TDT-measurements.
6.1.2 Reference Measurement
A
Reference Measurement
can be taken in addition to the TTC/12TC. The
software will move the zero point of the spatial axis (x = 0 cm) to the position
of the open-end (the coaxial cable or the TDR-probe tip).
47