10.
10.
10.
10.6
6
6
6 Water Calibration Procedures
Water Calibration Procedures
Water Calibration Procedures
Water Calibration Procedures
· The default unit for all factors in this program is millimeter so the figures displayed on the LCD
are the converted ones from the figures in millimeter.
-
-
-
- Pre
Pre
Pre
Pre-
-
-
-checking
checking
checking
checking
· Go to Menu 08 - 10.3.9
10.3.9
10.3.9
10.3.9 [[[[D
D
D
Density
ensity
ensity
ensity of
of
of
of the
the
the
the Lower Fluid
Lower Fluid
Lower Fluid
Lower Fluid Setting Mode]
Setting Mode]
Setting Mode]
Setting Mode] and check if the low density is
set to 1000 which is water density.
· If the density is set to the process condition, adjust the setting value using the Up / Down Keys.
· Checking of the low density setting is very important as this affects to the instrument accuracy a
lot.
· For example, if the calibration is conducted with Water (Density: 1000
ц
/
᎑
and if the low density
is set to 900, it will cause 10% errors.
· Be sure that the unit of setting values of this procedure is millimeter.
· Check if the installation and wiring have been done according to 7.5
7.5
7.5
7.5 [[[[Installation Procedure
Installation Procedure
Installation Procedure
Installation Procedure]]]] and
9.
9.
9.
9. [[[[Wiring
Wiring
Wiring
Wiring]]]].
1.
1.
1.
1. Zero Span (0%) Setting
Zero Span (0%) Setting
Zero Span (0%) Setting
Zero Span (0%) Setting
1) Fill water into the chamber or the tank till the 0% level.
2) Go to Menu 03 and set the value according to 10.3.4
10.3.4
10.3.4
10.3.4 [LRL Auto Calibration Mode]
[LRL Auto Calibration Mode]
[LRL Auto Calibration Mode]
[LRL Auto Calibration Mode] in this manual.
3) Go to Menu 01 to see if the value is set to “0”, and if not, change it to “0”.
4) Now the instrument will output 4
Ꮄ
when the level is 0mm. If the value would be set to 10 at
Menu 01, the instrument will output 4
Ꮄ
when the level is 10mm.
2.
2.
2.
2. Full Span (
Full Span (
Full Span (
Full Span (100%
100%
100%
100%) Setting
) Setting
) Setting
) Setting
1) Fill water into the chamber or the tank till the full span (100%) of the measuring range.
2) Go to Menu 04 and set the value according to 10.3.5
10.3.5
10.3.5
10.3.5 [L
[L
[L
[Length
ength
ength
ength Calibration Mode]
Calibration Mode]
Calibration Mode]
Calibration Mode] in this manual.
(The setting procedures must be finalized with the Enter Key to save the changes.)
Ex.1) If the actual length from 0% to 100% is 2345
р
and if the LCD PV Window shows ‘2300’, then
increase this value to 2345 using the Up Key. Finalize the setting procedures by pushing the
Enter key. In this example, 2345 will be the absolute standard value.
3) Go to Menu 02 and set the value according to 10.3.
10.3.
10.3.
10.3.3
3
3
3 [URV Calibration Mode]
[URV Calibration Mode]
[URV Calibration Mode]
[URV Calibration Mode] in this manual.
In the example above, the value should be set to 2345 so that the instrument outputs 20
Ꮄ
when
the water level is 2345
р
. If the value would be set to 2000, then the LCD PV Window shows 20
Ꮄ
when the water level is 2000. This means that [
Ꮄ
] only is changed without changing [
р
]. This
setting can adjust [
Ꮄ
] output by applying a random number using the URV Manual Calibration
function during the operation on demand.
3
3
3
3.
.
.
. Calibration
Calibration
Calibration
Calibration Verification
Verification
Verification
Verification
To see if the calibration has been done appropriately, change the water level in order of 0%, 20%,
40%, 60%, 80%, 100%, 80%, 60%, 40%, 20%, 0%, and check the outputs and the measuring values.
4
4
4
4.
.
.
. After completing all the proced
After completing all the proced
After completing all the proced
After completing all the procedures, change the Density of M
ures, change the Density of M
ures, change the Density of M
ures, change the Density of Menu
enu
enu
enu 08 to the current process’ density.
08 to the current process’ density.
08 to the current process’ density.
08 to the current process’ density.
CAUTION
CAUTION
CAUTION
CAUTION
M
M
M
Menu
enu
enu
enu 04
04
04
04 should be adjusted only during calibration procedure.
should be adjusted only during calibration procedure.
should be adjusted only during calibration procedure.
should be adjusted only during calibration procedure.
Adjusting during the normal operation will cause the change of the Span Gain.
Adjusting during the normal operation will cause the change of the Span Gain.
Adjusting during the normal operation will cause the change of the Span Gain.
Adjusting during the normal operation will cause the change of the Span Gain.