17. Performance and Technical Specifications
Compatibility
External device information
The IntelliBridge VueLink Open Interface (IVOI) Protocol is unidirectional, i.e. a monitor that adheres to
the protocol can display data received from the Tetra Graph but cannot remotely control the device.
Note that the user controls the monitoring using the TetraGraph Monitor graphical user interface when
interfacing with the Philips monitor.
The device transfers 9 measurement numerics and two waves to the monitor, using the Open Interface
protocol. The device makes use of the INOP (inoperable) functionality in the Open Interface protocol, to
alert the user about a loose connector, low battery or if somebody has pressed pause on the device.
Supported Devices
This integration supports the TetraGraph only and the minimum version that is supported is 34b.22b.12a.
Supported hosts
Senzime’s implementation of the IVOI protocol was carried out using Philips SpecTool (1040), the
IntelliBridge and VueLink Open Interface - Specification Tool (Version B). All hosts that support this
version of the IVOI protocol operating at 19200 Baud and open interface driver A.6 and higher will be
compatible.
IntelliVue monitor revision
Compatible Philips Monitors
(via IntelliBridge EC10)
Tetragraph is compatible with the following Philips Patient Monitors
when equipped with integrated or modular IntelliBridge EC10 Interface:
•
IntelliVue MP series (SW version H.15 or higher)
•
IntelliVue MX series (all SW versions)
Compatible Philips
Information Centers
(via IntelliBridge EC40/80)
Tetragraph is compatible with the following Philips Information Centers
when connected via EC40/80 hubs
•
PIIC iX IntelliVue Information Center iX (SW version B.0)
•
PIC iX Patient Information Center iX (SW version C.0 or
higher)
Compatible Driver
OpenInterface version A.6 or higher
How to find software revision
of the instrument
To locate the current version of the TetraGraph, please consult the IFU for the TetraGraph.
Also, the software version is shown on the display when the OFF button is pressed.
List of labels
In the tables below, all labels are provided with numeric codes from the Medical Device Interface Language (MDIL).
Labels
MDIL Text
ID
Label
Unit of
measure
Display
range
Definition
Description
0002-593c
EMG
%
0-120
"EMG"
Electromyography Low EMG
0002-f8ab
TOFcnt
Unitless
0-4
"TOFcnt"
Train of Four (TOF) count - Number of TOF
responses
0002-f897
TOFrat
%
0-120
"TOFrat"
Train of Four (TOF) ratio. Is the ratio between the
first and the fourth TOF response
0002-f88b
PTC
Unitless
0-20
"PTC"
Post Tetanic Count stimulation - PTC
0002-f8ac
Twitch
mV
0-50
"Twitch"
Twitch height of the 1Hz/0.1Hz stimulation
response
0002-f8a7
TOF1
mV
0-50
"TOF1"
Train of Four (TOF) first response value
0002-f8aa
TOF4
mV
0-50
"TOF4"
Train of Four (TOF) fourth response value
0002-f8a8
TOF2
mV
0-50
"TOF2"
Train of Four (TOF) second response value
0002-f8a9
TOF3
mV
0-50
"TOF3"
TrainOf Four (TOF) third response value