4.
Connect one 22 gauge wire from each valve to a neutral terminal labeled N on terminal block
J1 of the valve relay board. See above warning about using 3/8" cable with Sentry valve.
5.
The valve board in the standard Sample Sequencer comes pre-wired to the I/O Board. If using
the optional remote valve output board, then connect the 22 gauge seven conductor cable
(Belden 9430) from the color coded J1 terminals on the Sample Sequencer I/O board to color
coded J2 terminals on the valve relay board. Knockouts for 1/2" conduits are provided in
remote valve board enclosure. The maximum cable length recommended is 32 ft. (10m).
FIGURE 16 - Solenoid Sample Valve Connections