No.14S056-01
22
STC-CMC120APCL, STC-CMB120APCL
Specifications and Users guide
4.4 AOI Output Timing
This camera can set the AOI up to 8 different regions.
Variable Partial Start Line (VAS*), Variable Partial Effective Line(VAH*), Horizontal Start
Position(HAS*) and Horizontal Effective Pixel(HAW*) can be set for each eight regions. One region
can be set on horizontal direction. And the vertical area cannot overlap for each region.
An example of four areas of interest (AOI) are shown below:
AOI_
A
AOI_
B
AOI_
C
AOI_
D
VAH
A
A0-A1h
HAS
A
B0-B1h
VAS
A
90-91h
VAH
B
A2-A3h
VAS
B
92-93h
VAH
C
A4-A5h
VAS
C
94-95h
VAH
D
A6-A7h
VAS
D
96-97h
HAS
B
B2-B3h
HAS
C
B4-B5h
HAS
D
B6-B7h
4096
3072
HAW
A
C0-C1h
HAW
B
C2-C3h
HAW
C
C4-C5h
HAW
D
C6-C7h
Summary of Contents for CL Series
Page 1: ......