STEP 1 – UNPACKING THE UVT-LED

Make sure you have all the proper components. Refer 
to the packing list for the components you received 
based upon your order. 

Connect the power supply cable to a power adapter, 
then plug the power supply into a wall outlet to begin 
charging the battery. The battery requires approxi-
mately 15 hours to reach a full charge when plugged 
into a wall power outlet.

STEP 2 – CALIBRATING THE UNIT

Insert the UVT-LED into de-ionized water for initial cal-
ibration. Ensure it is inserted to level above measure-
ment area, and shake gently to remove air bubbles.

Short press the Read/Cal button to take a reference 
measurement. The reference measurement will show 
on the LCD display in approximately 10 seconds.

Press and hold the Read/Cal button for more than 3 
seconds until the LCD display backlight begins flash-
ing blue.

Once the LCD display reads measurement greater 
than 99.9%, the calibration procedure is complete. 

STEP 3 – TAKING A MEASUREMENT

Once calibrated, the UVT-LED is ready to take 
measurements. Please note that the unit should be 
re-calibrated every 100 measurements.

Before starting a reading, visually inspect the UVT-LED 
unit to ensure there is no material or debris on the 
unit that could contaminate the sample.

To begin a reading, place the UVT-LED into a sample. 
Make sure it is inserted to level above the measure-
ment area as shown.

Short press the Read/Cal button to take a measure-
ment. The measurement will show on the LCD display 
after approximately 10 seconds.

See UVT-LED manual for cleaning and maintenance information.