Sensorex SD7500CD Product Instruction Sheet Download Page 5

                                                                                                             

 PRODUCT INSTRUCTION SHEET 

Page 5 of 8

Intermittent Operation

Some facilities are only operated part of the time. When out 

of operation, electrodes must not be allowed to be exposed 

to air and become dry. Electrodes should be removed from 

such systems and stored in their bottles or caps or in a beaker 

filled, preferably, with pH 4.0 Buffer. In some instances, power 

to the meter is shut off; this condition can be harmful to elec-

trodes. Electrodes should be disconnected from un-powered 

meters.

Electrode Wiring

See the figures 10 through 17 to find wiring for your particu-

lar pH or ORP controller or transmitter.  

Figures 10-17 apply 

to SD7000CD and SD7000CD-ORP sensors only. 

  If you don't 

see a wiring diagram for your make and model number of 

controller, please contact Sensorex Technical support at tech-

[email protected] or call 714-895-4344.

Figure 18 is wiring for the SD7420CD & SD7420CD-ORP sen-

sors.  These sensors are loop powered 4-20mA.  12-30V DC 

power is required (>24V DC is recommended). 

Figure 19 is wiring for SD7500CD & SD7500CD-ORP sensors.

These sensors DO NOT REQUIRE POWER.  Do not connect to 

power.  These are self-powered.  

Aquametrix 

Shark 

Controller

Aquametrix 

Shark 

TX/P

Transmitter

 FIG. 10

 FIG. 11

GLI PRO P3

 FIG. 12

Form:  InstructSD7000CD-06 [Rev: 2016-10-10] 

Summary of Contents for SD7500CD

Page 1: ...trode by fiscal week and year Example 4613 week 2 of November 2013 3 After exposure to a sample buffer or rinse solution carryover can be minimized by blotting never by wiping the elec trode with a cl...

Page 2: ...electrode span errors but will not affect the speed of response 8 Salt bridge should be replaced when the electrode readings cannot be corrected by the meter s controls and or when their speed of res...

Page 3: ...ess above 7 pH it is advisable to soak it 4 5 minutes in a weak acid solution hydrochloric recommended Place the sensor back in the soap solution for 2 3 minutes Rinse it in clean warm water Hard Coat...

Page 4: ...rated but quinhydrone based solution or other standards can be used to verify ORP sensor performance See FIG 9 Two Point Calibrations Two point calibrations correct for both the pH electrode s offset...

Page 5: ...gh 17 to find wiring for your particu lar pH or ORP controller or transmitter Figures 10 17 apply to SD7000CD and SD7000CD ORP sensors only If you don t see a wiring diagram for your make and model nu...

Page 6: ...cted the same way rinse the sensor with water and put it in a pH 4 or pH 10 buffer solution Wait for the temperature of the sensor and buffer to reach room temperature 6 Compare the mV reading in the...

Page 7: ...V Temp Pressure Min 32 F 0 C Max 185 F 85 C Speed of Response 95 in 5 seconds Wetted Materials PPS Body PVDF Salt bridge junction pH glass pH Platinum ORP Titanium ground pin Viton o rings Cable Lengt...

Page 8: ...or ORP Green Reference Black Yellow Temperature PT1000 RTD Ordering Information Part Number Description SD7000CD Differential pH sensor with replaceable salt bridge 20ft cable tinned leads 1 5 NPT mo...

Reviews: