Congratulations on your purchase of your Sensorex SD7000 

Series pH or ORP true differential electrode.  Please follow 

instructions and tips to get the maximum life out of your 

electrode.

Introduction

Electrode Care and Use Tips

1. The SD7000 Series pH and ORP electrodes are shipped with a  

    cap containing a solution of pH 4 buffer and potassium  

     

    chloride.  

    The cap should remain on the electrode until it is used. If the 

    electrode is used infrequently the cap and its solution should 

    be saved and the electrode stored in it. Store pH and ORP 

    electrodes in pH 4 buffer if soaking solution is discarded. 

    Before using the electrode remove the soaker cap as shown     

   in FIG 1.

2. Sensorex true differential electrodes have a 2 year prorated           

warranty from the date etched in the electrode body as shown         

   in FIG 2.   Electrodes in inventory should be rotated so that the      

    older electrodes are used first. The date code is stamped on 

each electrode by fiscal week and year (Example: 4613 = week 2    

of November 2013). 

3. After exposure to a sample, buffer or rinse solution, carryover 

    can be minimized by blotting—never by wiping—the elec-

    trode with a clean non-abrasive paper or a clean cloth towel. 

    

Do not use a brush on pH glass. 

4. As a rinse solution, measure and use a portion of the next 

    sample or buffer. This action will minimize carryover 

    contamination. 

5. When calibrating pH electrodes, use a buffer close in value to 

    that expected from the sample for one-point calibrations or as 

    the second buffer for two-point calibrations (See below). This 

    action will minimize span errors. ORP electrodes are typically 

    not calibrated. However, a solution to verify electrode perfor-

    mance should be used periodically. 225mV solution, Zobell's 

    solution (Part no. B225), is a good choice for single point 
    verification.

                                                                                                             

 PRODUCT INSTRUCTION SHEET 

Parts covered by this product data sheet include: 

SD7000CD, SD7000CD-ORP, SD7500CD, SD7500CD-ORP, SD7420CD, SD7420CD-ORP

    SD7000 Series Differential 
  pH/ORP Electrode Instructions

 FIG. 1

 FIG. 2

Form:  InstructSD7000CD-06 [Rev: 2016-10-10] 

 

Mo

de

l S

D7

00

0C

D D

iffe

ren

tia

l p

H

4615

Summary of Contents for SD7500CD

Page 1: ...trode by fiscal week and year Example 4613 week 2 of November 2013 3 After exposure to a sample buffer or rinse solution carryover can be minimized by blotting never by wiping the elec trode with a cl...

Page 2: ...electrode span errors but will not affect the speed of response 8 Salt bridge should be replaced when the electrode readings cannot be corrected by the meter s controls and or when their speed of res...

Page 3: ...ess above 7 pH it is advisable to soak it 4 5 minutes in a weak acid solution hydrochloric recommended Place the sensor back in the soap solution for 2 3 minutes Rinse it in clean warm water Hard Coat...

Page 4: ...rated but quinhydrone based solution or other standards can be used to verify ORP sensor performance See FIG 9 Two Point Calibrations Two point calibrations correct for both the pH electrode s offset...

Page 5: ...gh 17 to find wiring for your particu lar pH or ORP controller or transmitter Figures 10 17 apply to SD7000CD and SD7000CD ORP sensors only If you don t see a wiring diagram for your make and model nu...

Page 6: ...cted the same way rinse the sensor with water and put it in a pH 4 or pH 10 buffer solution Wait for the temperature of the sensor and buffer to reach room temperature 6 Compare the mV reading in the...

Page 7: ...V Temp Pressure Min 32 F 0 C Max 185 F 85 C Speed of Response 95 in 5 seconds Wetted Materials PPS Body PVDF Salt bridge junction pH glass pH Platinum ORP Titanium ground pin Viton o rings Cable Lengt...

Page 8: ...or ORP Green Reference Black Yellow Temperature PT1000 RTD Ordering Information Part Number Description SD7000CD Differential pH sensor with replaceable salt bridge 20ft cable tinned leads 1 5 NPT mo...

Reviews: