Recording

Temperature data

Displayed Information

Stopped recording

Limits exceeded

Marked point

To mark an important event (“Date Stamp”)

•  TempTale4 Probeless Dry Ice monitors provide an option to “Date Stamp” or mark an
  important event at any time during the monitoring cycle that will be visible when viewing
  the data. The mark is displayed as an arrow when viewing the Graph and shown in italics
  when viewing Tabular data.  
•  To mark an event while the monitor is recording, press and release the “Start” button.
•  The “arrow” icon will appear temporarily in the lower left corner and the temperature history
  will appear on the display. Data appears as follows:

    1.  Average temperature
    2.  Highest temperature recorded
    3.  Total time above high temperature limit
    4.  Lowest temperature recorded
    5.  Total time below low temperature limit

Alarm

When the TempTale4 Probeless Dry Ice monitor  is exposed to a temperature outside a pre-
programmed limit and time threshold, an “Alarm bell”        will appear in the lower right corner 
of the LCD display.  

Stopping a TempTale4 Monitor

Press and hold (1 to 3 seconds) the Stop button until the “Stop sign”        icon appears in the up-
per right corner of the LCD display. 

Preparing a TempTale

®

4 Probeless Dry Ice Monitor for Shipment 

•  Prepare your container for shipment according to your SOP
  (Standard Operating Procedure) for dry ice pack outs.
•  At the point in the SOP where dry ice fills a portion of the box
  and the product is enclosed, press and release the Green/left Start
  button on the TempTale4 Probeless Dry Ice Monitor.
•  Verify that the “Sunshine”       icon appears in the upper left
  corner of the display.  
•  Place the TempTale4 Probeless Dry Ice Monitor as close to your
  product as possible.
•  Complete the SOP by continuing to add the desired weight of dry
  ice prior to closing and labeling the shipping container.
•  Now your dry ice shipment is ready for shipping.
•  If your monitor has been programmed with a Start-up delay the
  monitor will not begin recording data until the Start-up delay has passed.

Receiving a TempTale 4 Probeless Dry Ice Monitor

• 

Important:

 Always use protective gloves and caution when removing the TempTale4

  Probeless Dry Ice monitor from the –80°C dry ice environment.
•  When the dry ice shipment arrives, the Receiver should remove the TempTale4 Probeless
  Dry Ice monitor and press the Red/right Stop button to manually stop the unit.

1

  

•  Verify that the stop sign icon appears on the LCD panel indicating the TempTale4 Dry Ice
  monitor has been stopped.

2, 3, 4

Notes:

  

1.  If monitor is not stopped manually, the monitor will continue to run and record time and temperature data until
 

the monitor is downloaded or until the pre-programmed trip length is reached.

2. The LCD of the TempTale4 Probeless Dry Ice monitor requires a ‘warm-up’ period of up to 5 minutes at room
 

temperature to display summary data and icons after exposure to extremely low temperatures. 

3.  After removing the TempTale4 Probeless Dry Ice monitor from the container, the monitor can be downloaded to
 

your PC immediately if you have TempTale Manager

®

 2.4 or TempTale Manager Desktop installed on your PC.

 

a)  If you are running TempTale Manager 2.4, the Monitor Type displayed in Graph, Summary and Tabular

 

 

Views is Ambient instead of TempTale4: Dry Ice. 

 

b)  If you are running TempTale Manager Desktop, the Monitor Type displayed in Graph, Summary and 

 

 

Tabular Views is TempTale4: Dry Ice.

4.  Download the TempTale4 Probeless Dry Ice monitor to your PC using the Interface Plus Reader by placing the
 

monitor in the Interface plus with the LEDs aligned.