VALVE CONFIGURATION
Seven out of the eight valve inlet connections on the back of the
SMART CAL 360
are fully configurable (valve 6 is
reserved for fresh air). By default, valves will be setup for the most common configurations:
Valve 1: 2.5%V/V Methane / 100ppm CO combo gas
Valve 2: Line Gas (for 100% V/V Methane calibration)
Valve 3: 25ppm H2S
Valve 4: 0.1% V/V (1000ppm) Methane
Valve 5: 1.1% V/V Propane
Valve 6: Clean air (not configurable)
Valve 7: Clean air
Valve 8: 100ppm Ethane
If any of your instruments require calibration gas that isn’t listed above, or if
you need to change the existing configuration, configure the valves by
following the instructions below.
1. From the main display, select the menu button on the bottom right.
2. Enter the supervisor password (contact SENSIT if you need the
password).
“Supervisor Menu” should appear at the top of the screen.
3.
Locate and select “Valve Configuration” from the list.
4. For each gas you want to setup, select one of the valves from the top of
the list, whether you will be using a demand or positive pressure
regulator, and the appropriate gas from the list.
5. After all the valves are configured, hit the save icon on the bottom right.
Hit the arrow in the upper left to return to the main display.