16
MAINTENANCE AND CARE
To keep the
SMART-CAL 360
clean, and to protect and extend the life of the touchscreen, it is recommended to
keep the provided dust cover on the station when it is not in use.
It is recommended to turn all regulators and/or gas valves off when the
SMART-CAL 360
is not in use.
If your calibration setup has line gas plumbed directly into the station (instead of bottled 100% volume methane), it
is recommended to use an on-demand regulator.
F
ILTER
R
EPLACEMENT
The
SMART-CAL 360
incorporates a hydrophobic dirt and water filter on each of the 8 hose connections on the
back of the station. Over time is it possible these filters may become clogged with debris or oils from inside the gas
cylinders
, which would result in “Check Gas Supply” errors. If this happens, or if the filter looks damaged or frayed,
the filter will need to be replaced.
To replace a filter:
1. Disconnect the tubing from the fitting and remove the old filter using a
9/16” (15mm) wrench or socket. The plastic housing and filter are one
piece and will be replaced together.
2. Discard the old O-ring and place the new one in the cavity as pictured.
3. Install the new filter (do not overtighten) and reconnect the tubing.
4. Test that the filter is sealed and working properly by performing either a bump test or calibration using any
compatible instrument that would utilize that connection.