SCAL-100

QUICK-START INSTRUCTIONS

1. Find suitable location in gas-free area Desk or 

wall mount is possible.

2. Connect gas supply at 20psi inlet pressure. 

3. Attach power supply (unit will turn on 

automatically).

4. Attach to LAN (Scal-N software installed 

separately)

5. Turn on or zero P100

6. Place P100 into holder and “snap” in place.  

Top in first and bottom second.

Gas supply will indicate green light if adequate 

gas pressure is sensed.

“Busy”  LED  will  flash  during  operation  and 

upload.

Bump  Test  will  be  attempted  first  for  no  more 

than 30 seconds.

If passed then PASS indicator will flash.

If failed calibration will automatically be 

attempted.

Calibration time is 2 min max.

PASS or FAIL indication upon completion.

7. Remove Detector by releasing bottom first.

Data is stored for later upload through server 

operation.