EXPERT MENU

To access the expert menu from the work display (READY FOR MEASURENT) press the release the “C” 

button until EXPERT MENU is displayed.  This menu is password protected. The “C” Button will increase and 

“B” Button will decrease the number.  Once password is set press “A” Button to enter the menu. 

From this menu the following features can be viewed by pressing the “C” button to scroll through the options. 

Enter the selection by pressing the “B” button and adjusting by using the “C” button or the onscreen prompts.  

Pressing the “A” button repeated will return you to the working display.

No.   Name 

Description

CONTRAST 

Set contrast of display

2. 

LANGUAGE 

Allow user to select language of the system

3. 

GET C-CONST 

This process finds the constant value with the internal methane cell.

 

 

Perform this whenever you refill or replace the cell. 

4. 

CAL INTERVAL 

Set how often calibration should be performed.

5. 

BUMP INTERVAL 

Set how often bump test should be performed.

6. 

BUMP PPM-C2H6 

Allow you to set what PPM you would like to use for manual bump test.

7. 

BUMP LO LIMIT 

This is in %. Define minimum bump test response required to pass test.

8. 

BUMP HI LIMIT 

This is in %. Define maximum bump test response required to pass test.

9. 

ALARM PPB 

This is not implemented.

10. 

ALARM PPM 

This is not implemented.

11.   PURGE TIME 

Time in Sec unit should purge system before shutting down.

12. 

ERASE BUMP LOG  Allows you to erase bump logs.

13. 

ERASE CAL LOG 

Allows you to erase calibration logs.

14. 

ERASE MEAS LOG  Allows you to erase measurement logs.

USER MENU

To access the menu press and release the “C” button.  From this menu the following features can be viewed 

by pressing the “C” button to scroll through the options. Enter the selection by pressing the “B” button and 

adjusting by using the “C” button or the onscreen prompts.  Pressing the “A” button repeatedly will return you to 

the working display.

No.   Name 

Description

WARNINGS 

1. SD Card not inserted

 

 

2. Blue Tooth not communicating

 

 

3. GPS not communicating

 

 

4. BUMP Test skipped

 

 

5. Calibration Skipped

2. 

SET TIME 

Allow user to set time. 

3. 

SET DATE 

Allow user to set date.

4. 

BUMP TEST 

Allow user to perform Bump Test

5. 

CALIBRATION 

Allow user to perform Calibration

6. 

VIEW BUMP LOG 

Allow user to scroll and view bump log including autobump log.

7. 

VIEW CAL LOG 

Allow user to scroll and view calibration log including skipped log.

8. 

VIEW MEAS LOG 

Allow user to view measurement log

9. 

ERASE MEAS LOG  Allow user to erase measurement log. This will erase all measurement logs.

10. 

VIEW SETTINGS 

Allow user to view settings of device.

11.   EXPERT MENU 

Provide access to expert menu

12. 

FACOTRY MENU 

Provide access to factory menu

13. 

TEST MENU 

Provide access to test menu

Summary of Contents for IRED

Page 1: ...instructions before use Patented SENSIT IRED INFRARED ETHANE DETECTOR 851 Transport Drive Valparaiso IN 46383 8432 MADE IN USA SENSIT Technologies is in compliance with ISO 9001 2008 Phone 888 4SENSI...

Page 2: ...to use Always start the IRED in an area known to be gas free Tampering with this product may void the warranty Use only Sensit Technologies approved parts and accessories Never use an instrument known...

Page 3: ...cription 4 Parts and Accessories 5 Product Specifications 6 Product Features 6 Electronic Features 7 Display Features 8 Housing Features 9 Filters 10 Field Operations 11 12 Bump Test 13 Calibration 14...

Page 4: ...an output that is changed when certain gases absorb the light output The filter only allows specific light wavelengths to be monitored and measured The concentration of gas is proportional to the amo...

Page 5: ...Part 873 00024 Survey Probe Filter Blue 1 Filter Part 360 00064 SENSIT IRED Calibration Kit IRED Calibration Kit Part 881 00100 Kit includes 1 Regulator 1 Cal Adapter 1 100pm Ethane Nitrogen 58L SENS...

Page 6: ...ic Sensor Specifications IR Sensor Method Advanced Infrared Absorption Spectroscopy Detection Range 250ppb 1000ppm Warm up 30 min Calibration required 3 months Product Features The IRED is constructed...

Page 7: ...ghting IR communication port LCD Button Operation The Sensit IRED is operated with the use of three buttons below the display They are labeled A B C The A button Activates Deactivates the instrument A...

Page 8: ...once for manual data save GPS Connected GPS enabled not connected 8 Display Features The display provides all information including Gas concentration in PPB and PPM Battery voltage status Pump Status...

Page 9: ...taining screws Charge Complete Green Charging Red Status Yellow Charger Connection The battery pack has LED s to indicate charge status Charging should only be performed in an area know to be free of...

Page 10: ...r shoulder strap attachment Clips on the strap will attach securely Hydrophobic Filter Assembly WARNING Do not operate without proper filter Damage may occur to the pump and other internal parts Only...

Page 11: ...nd Time d Warm up is indicated PLEASE WAIT with a 600 second countdown timer The countdown timer will begin after an initial warm up of 5 10 minutes e Calibration due date indicating when calibration...

Page 12: ...imer c The measuring of ethane is indicated by the MEASUREMENT heading and the ppb ppm readings During a measurement 4 consecutive 10 second readings are taken Each reading is averaged over the 10 sec...

Page 13: ...stable reading must occur before the process automatically continues 5 If the reading is not stable the zeroing process will automatically re initiate indicated by a RE ZEROING message This process r...

Page 14: ...stable reading must occur before the process automatically continues 5 If the reading is not stable the zeroing process will automatically re initiate indicated by a RE ZEROING message This process r...

Page 15: ...mplemented 10 ALARM PPM This is not implemented 11 PURGE TIME Time in Sec unit should purge system before shutting down 12 ERASE BUMP LOG Allows you to erase bump logs 13 ERASE CAL LOG Allows you to e...