BUTTON (B) once to enter the menu, press & release the SAVE/
ZERO BUTTON (C) to scroll until the bottom line reads CAL
LOG. Prepare the optional IR printer. Aim the IR window on the
right side of the instrument at the IR printer. Press & release the
BH TEST/MENU BUTTON (B) to print the log. To exit this menu,
press & release the POWER/MUTE BUTTON (A) to return to
working display.
SHOW / PRINT a SESSION LOG
To Show a Session Log
From the working display, press & hold the BH TEST/MENU
BUTTON (B) until the top line reads USER MENU and the bot-
tom line reads PRINT MENU. Press & release the SAVE/ZERO
BUTTON (C) to scroll until the bottom line reads SHOW SES
LOG. Press & release the BH TEST/MENU BUTTON (B) once
to enter the menu. SESSION 1 will be displayed. This is the
most recent data saved. Press & release the SAVE/ZERO BUT-
TON (C) to scroll to the session number you want to view. Press
& release the BH TEST/MENU BUTTON (B) once to enter that
session’s recorded day, month and time. Press & release the
BH TEST/MENU BUTTON (B) once again to access the gas
readings for that session.
Press & release the SAVE/ZERO BUTTON (C) to scroll to view
the individual gas readings saved. Press & release the POWER/
MUTE BUTTON (A) once to access a different session number.
Press & release the SAVE/ZERO BUTTON (C) to scroll to a
new session number.
The standard number of available stored sessions is factory set
at 6 but is factory adjustable up to 100. To exit this menu, press
& release the POWER/MUTE BUTTON (A) to return to the working
display.
To Print a Session Log
From the working display, press & hold the BH TEST/MENU
BUTTON (B) until the top line reads USER MENU and the bot-
tom line reads PRINT MENU.
Press & release the BH TEST/MENU BUTTON (B) once to en-
ter this menu, SESSION LOG will be displayed.
Prepare the optional IR printer. Aim the IR window, on the right
side of the instrument, at the IR printer. Press & release the BH
TEST/MENU BUTTON (B) to print the log.
Press & release the POWER/MUTE BUTTON (A) to return to
the working display.
BUMP TEST
From the working display, press & hold the BH TEST/MENU
BUTTON (B) until the top line reads USER MENU and bottom
line reads PRINT MENU. Press & release the SAVE/ZERO BUT-
TON (C) to scroll until the bottom line reads BUMP TEST.
Prepare the appropriate certified gas mixture for your instrument
model (see proper gas mixtures listed in the Calibration sec-
tion).
Apply the gas to the instrument and press & release the BH
TEST/MENU BUTTON (B) to start the BUMP TEST.