SenseAir pSense II Short Instruction Manual Download Page 2

AUTO POWER OFF

Fig.14

The meter turns off automatically after 20 minutes of inactivity.
To override the function, first, the meter is in power off status.
Hold down         and       for two seconds to turn on the meter 
until “

n

” appears (Fig.14). 

SET

ON/OFF

HOLD

NOTE:

 

Auto sleep function will be disabled during calibration &
auto recording mode.  Auto sleep function WILL NOT be 
disabled automatically via using adaptor. 

The sensor has an automatic calibration function to secure long-term accuracy. If a 
calibration is desired, press                     simultaneously for 3 seconds to turn on the 
meter and enter calibration mode. Press        or        key to select 400ppm CO

2

 

calibration (Outside Air calibration). Press        to start and see“CAL” and CO

2

 value 

blinking on LCD. Wait about 10 minutes until the blinking stops and the calibration is 
completed and back to normal mode. While choose 0ppm CO

2

 calibration, remove 

the cap from meter rear side and plug the CO

2

 CALIBRATOR pipe into the sensor 

hole beneath. Wait about 10 min until the blinking stops and the calibration is 
completed and back to normal mode.

MAX/AVG

Humidity Calibration

This meter can be calibrated either via 35% & 80% salt bottle or humidity generator.

o

The ambient condition is recommended as stable 25 C. While the meter is powered
off, plug the sensor probe into 35% salt. Press       ,       ,        simultaneously for three
seconds to turn on the meter and enter calibration mode. Press        or        to select
35.0% calibration and press        to start. “CAL” and calibrating value are blinking on
the LCD. 

CAUTION:
Do not calibrate the humidity without the default calibration salt. Otherwise,
it will cause permanent damage. Contact SenseAir for calibration salt or 
services.

 

Single point calibration will cause error code E11 or non-accurate reading. 
Always do dual point calibration to complete a process. Suggest to calibrate 
low humidity first. 

CO2 Calibration 

CAUTION: 
Do not calibrate the meter in the air with unknown CO  concentration. 

2

Otherwise, it leads to inaccurate measurements.  
Ensure the batteries are with fully charged during the calibration to prevent 
frominterruption or failed calibration

When the accuracy becomes a concern 

after long time usage, return to dealers for standard calibration.

Technical Data

Humidity/accuracy:                  

Trouble Shooting

Error    Messages                                    Solution    

E01                 
                             
E33                  
E02      
E

 

03      

E

 

04       

             
E11                           
E

 

31             

E32                 
E33               
   

Temperature/accuracy:           

CO

2

 /accuracy:                          

Storage and transit:                  

Operating limit for CO

2

:           

Memory:                                     

Dimensions (LxWxH):              

Weight:                                       

Battery type:                              

DC power type:                         

PC connection:                         

CO

2

 CALIBRATOR:                          00-0-0022  Zero Calibration Kit

Optional Accessor y

Art.No.

Batter y low indicator

A battery low icon “ 

   

“ appears when the supplied power is too low to maintain a

functional operation. Please plug in the 9VDC adaptor enclosed to charge the batteries.
While charging the batteries, you still can power on the device and operate as usual.

SET

ON/OFF

START

ESC

MEM

HOLD

MEM

MIN

When meter is powered on , hold down         for more than two seconds until entering 
meter setup mode. To exit setup, press        to return to normal mode.

Press       or       key to select the program and press        key to select. 
The programmable setting flashes on display, press          or        key to select and 
press           key to confirm. To leave without saving, press         key to return. 

           

  

DATA LOGGING

The meter can automatically record HUMIDITY/TEMP/CO

2

 for long term environment 

monitor. The memory capacity is 6000 points for each parameter. Users can set up 
sampling rate from one second to 4hours, 59 minutes and 59 seconds. The factory 
preset rate is 30 second.

START
   ESC

REC

Fig.19

CO

2

REC

AM

Repeat above to start another run of logging. While the logging 
space is full, it shows “FULL”. To terminate data logging, press        
         for two seconds,         -icon stops flashing. 

Press        key to switch to the mode you want to see. Except the 
mode switching function, auto power off, MIN/MAX/AVG, 
manual record, hold and recall functions are all disabled during 
logging. 

REC

START
   ESC

MODE

RECALL

Meter Setup

SET

ON/OFF

START

 ESC

HOLD

MEM

MAX/AVG

MIN

HOLD

MEM

MAX/AVG

MIN

START
   ESC

P10, 99 points MEMORY clear

P11, choose yes or no

Note:

“        “is displayed while 
memory is cleared.

P20, auto LOG data clear

P21, choose yes or no

Note:

“        “is displayed while 
memory is cleared.

P30, CO2 ALARM setup

P31, choose from 100 
         to 9900

Note:

-Factory preset 1400ppm.
-Adjustable scale is every
 100ppm

o

o

P51, choose 

or

  F 

 

Note:

o

-Factory preset  C 

P50, Temp. UNIT setup

o

C

o

F

P61, choose from 
         

04:59:59 to 

           00:00:01

Note:

-Factory preset at 30 sec.
-The format is 
  hour:minute:second 

P60, logging sampling RATE

P71, choose from
         700 to 1990hpa 
       

Note:

-Factory preset at 1013hpa
-The adjustable scale is
  every 1hPa.

P70, PRESSURE compensation

MANUALLY RECORDING

REC

MEM

HOLD

MEMORY RECALL

In

 

normal or hold mode, press         key for more than two seconds until the           -icon 

flashes. Press        

 

or        to scroll the memories. The memory sequential number 

displays on main LCD first and reading comes after (Fig.16 &17). Press        

 

key to 

switch the display parameter (Fig.18).The time displayed in memory recall mode is 
the recording time of this entry.

 

Recall

HOLD

MEM

Fig.15

REC

AM

MODE

RECALL

Fig.16

Recall

Fig.17

CO

2

Recall

Fig.18

Recall

%rh

MODE
RECALL

After sampling rate is selected, press          for two seconds under normal mode to start 
logging. The         -icon flashes to indicate the logging status and LCD main display 
shows the real time value. Lower displays are real time clock (Fig.19). 

To setup the sampling rate, when meter is powered on , hold down         for more than 
two seconds until entering meter setup mode. Press       or       key to select the 
program P60 (rAtE) and press        key to enter.        

Press        or       to select the sample rate from hour, minute to second and then press
          to confirm. Press         to return to normal. 
           

SET

ON/OFF

MAX/AVG

MIN

START

ESC

HOLD

MEM

MAX/AVG

MIN

HOLD

MEM

The meter features a 99 point memory. In the normal or hold mode, press        key 
to start recording, the          -icon and the main display will be flashed three times. The 
memory sequential number will show on the main display with 99 points in maximum. 
The data in the memory contains all parameters information and not only the displa-
yed parameter. When all 99 points memory are occupied, a “FULL” message will be 
displayed on the screen. If you think the reading is changing too fast, you can press 
       key to freeze current reading first then manually record the data. This recording 
function will be disabled in the MIN/MAX/AVG mode.

P40, CO

2

 ABC setup

P41, Choose on or off

 

 

Note:

-Factory preset ABC 
  function as ON

SET

ON/OFF

START

ESC

MEM

HOLD

MEM

MAX/AVG

MIN

® 

SenseAir AB

Stationsgatan 12
Box 96
820 60 Delsbo
Phone: +46(0)653-71 77 70 
Fax: +46-(0)653-71 77 89 
 

0.1% to 99.9%/ ±3.0%(10~95%@25 °C), ±5% (other)

o

o

o

-20 to 60 C/+0.3 C@5~40 C

0 to 9(30ppm+5% of reading)@0~5000ppm

o

-20 to 60 C/10 to 90%rh, non-condensing.

o

o

0 to 50 C, non-condensing / Electronics: -20 to 60 C

99 manual records, 18000 data logging records.

234x77x42 [mm]

200g

Four pcs AA re-chargable batteries

9VDC, 500mA (Standard accessory).
mini USB port

Turn off meter and re-start again
or re-charge the battery
Retry CO  calibration

2

Put meter in normal condition  
Put meter in normal condition
Solve the original data error

Retry humidity calibration
Return for repair
Return for repair
Return for repair

CO2 sensor is out of order
or the battery is too low 
CO  sensor is out of order 

2

measured value is under range     
measured value is over range       
Original data error results in
this error (DP, WB)
RH calibration error  
Temp. sensor or AD damaged
Memory IC damaged          
RH sensor or circuit damaged

P81, choose 12 or 24
          hour format 
       

Note:

-Factory preset at 12H & 
 dd-mm-year format.

P80, REAL TIME CLOCK setup

         choose DATE 
          format 
       

         Input Year/
         Month/ Day
         
       

year
month
date

         Input Hour/
         Min./Sec.
         
       

Hour
Minute
Second

or