<<    Operating Manual for Inhaled Gas Detector

 

 

 

 

26 

©

Senko 2019. All Rights Reserved Proprietary and Confidential to Senko 

 

4000X Register Read 

구분

 

Address 

Bits 

Description 

Monitoring Status 

40001 

BIT0~3 

0 : Warmup 
1 : Measure Mode 
2 : Inhibit Alarm 
3 : Inhibit Alarm/Fault 
4 : Inhibit Full 
5 : Reserved 
6 : Test Mode 
7 : 4-20mA Calibration Mode 
8 : Flow Calibration Mode 
9-15 : Reserved 

BIT4 

Fault Active Status 

BIT5 

Reserved 

BIT6 

Alarm 1 Active 

BIT7 

Alarm 2 Active 

BIT8 

Alarm 1 Relay energized 

BIT9 

Alarm 2 Relay energized 

BIT10 

Fault Relay energized 

BIT11 

Heartbeat Bit(2

 

간격

  Toggle) 

BIT12~15  Reserved 

Cartridge Selection 

40002 

BIT0~7 

Gas ID(Sensor Type) 

BIT8~15 

Reserved 

측정

 

가스

 

농도

(

실수

40003 

BIT0~15 

실수형

 

가스

 

농도

 

측정값

(

상위

  2byte) 

40004 

BIT0~15 

실수형

 

가스

 

농도

 

측정값

(

하위

  2byte) 

측정

 

가스

 

농도

(

정수

)  40005 

BIT0~15 

정수형

 

가스

 

농도

 

측정값

 

Fault Code 

40006 

BIT0~15 

Fault Code 

Decimal  Point  and 
Units 

40007 

BIT0~2 

Decimal Point Indicator(0~3) 

BIT3~7 

Reserved 

BIT8~15 

1 : ppm(

농도

 

단위

2 : ppb(

농도

 

단위

3 : % volume(

농도

 

단위

4 : %LEL(

농도

 

단위

16 : mA 

온도

 

측정값

 

40008 

BIT0~15 

온도

 

측정값

(Signed 16bit Integer) 

Reserved 

40009 

BIT0~15 

Reserved 

Reserved 

40010 

BIT0~15 

Reserved 

Flowrate 

40011 

BIT0~15 

Flowrate(cc/min) 

Reserved 

40012 

BIT0~15 

Reserved 

1

  Alarm 

설정

 

(

실수

40013 

BIT0~15 

실수형

  Alarm 1 

설정

 

(

상위

  2byte) 

40014 

BIT0~15 

실수형

  Alarm 1 

설정

 

(

하위

  2byte) 

Summary of Contents for SI-H100

Page 1: ......

Page 2: ...anual 1 3 2 4 3 Overview 5 4 6 5 7 1 7 2 8 3 8 4 9 5 9 6 able Power board 10 7 Dimensions 11 11 8 Key 12 1 Key Description 12 2 Key State 12 9 Power On 13 10 Menu 14 1 Normal Measurement Menu 15 2 Set...

Page 3: ...2 Senko 2019 All Rights Reserved Proprietary and Confidential to Senko 9 Test mode 24 11 MODBUS Address Map 25 1 RS485 Interface setting 25 2 TCP Interface setting 25 3 MODBUS RS485 TCP Register 25 1...

Page 4: ...Operating Manual for Inhaled Gas Detector 3 Senko 2019 All Rights Reserved Proprietary and Confidential to Senko 1 SI H100 SI H100 FAULT 4 4 20mA 3 MODBUS TCP LAN PoE WARNING...

Page 5: ...Operating Manual for Inhaled Gas Detector 4 Senko 2019 All Rights Reserved Proprietary and Confidential to Senko 2 SI H100 4 NF3 Pyrolyzer...

Page 6: ...Operating Manual for Inhaled Gas Detector 5 Senko 2019 All Rights Reserved Proprietary and Confidential to Senko 3 Overview Figure 1 Overview...

Page 7: ...Senko 4 124mm H X 65mm W X 155mm D 2Kg DC 24V 10 PoE 36V 57V Typical 48V 500mL min MAX 900 5 0W Graphic LCD 160 X 100 Relay 1 Alarm 2 Alarm Fault Alarm Analog 4 20mA Digital RS 485 TCP Ethernet 30m FE...

Page 8: ...9 All Rights Reserved Proprietary and Confidential to Senko 5 1 Power LED Green Alarm1 LED Red Alarm2 LED Red Fault LED Red LCD Menu button Down button UP button Select button Sensor cartridge window...

Page 9: ...Operating Manual for Inhaled Gas Detector 8 Senko 2019 All Rights Reserved Proprietary and Confidential to Senko 2 3 Unit cover Mount bracket hole cover...

Page 10: ...ights Reserved Proprietary and Confidential to Senko 4 5 Ethernet PoE socket Unit cover Sensor cartridge Mounting bracket assembly Main sampling assembly Main sampling assembly bracket assembly fixed...

Page 11: ...Senko 6 able Power board Power ON OFF switch Ethernet PoE socket PoE PoE Power cable DC 24V Power cable Ground ALARM LOW 2 Relay ALARM HIGH 2 Relay Fault Relay RS 485 A RS 485 B 4to20mA out Signal 4to...

Page 12: ...Operating Manual for Inhaled Gas Detector 11 Senko 2019 All Rights Reserved Proprietary and Confidential to Senko 7 Dimensions...

Page 13: ...d Proprietary and Confidential to Senko 8 Key 1 Key Description Key Name Description Menu Menu Up List Focus Up Long Setting Focus Down List Focus Down Long Setting Focus Select 2 Key State State Pres...

Page 14: ...Operating Manual for Inhaled Gas Detector 13 Senko 2019 All Rights Reserved Proprietary and Confidential to Senko 9 Power On Power LED Green Version Measure 15sec Booting Warm up Measure...

Page 15: ...Gas Detector 14 Senko 2019 All Rights Reserved Proprietary and Confidential to Senko 10 Menu Menu Normal SET 1 SET 2 SET 3 CAL Normal Select Review Menu Review Normal Power On Normal Measure SET 1 SE...

Page 16: ...nko 1 Normal Measurement Menu 24 Gas Type O3 Gas Type ppm Gas Sensor On Off State Pump 2 Setup Calibration Review List 1 Menu SET1 SET2 SET3 CAL REVIEW 2 Up Down List Focus 3 Select 4 Menu Menu 3 Setu...

Page 17: ...Calibration 4 Pump Freq Gas Pump Frequency 5 Pass Code Setup Calibration 6 Inhibit Inhibit 7 Buzzer Alarm Buzzer On Off 8 Resp Factor Gas 9 Sensor Off Sensor On Off VOC 10 Hidden Area Gas Range Hidde...

Page 18: ...Long Focus Up Long Focus Down Long Focus Menu Select 2 Pass Code Pass Code 0000 Menu Pass Code Default 0000 Inhibit Up Down Inhibit Menu Select None Inhibit Alm Alarm Inhibit Alm Flt Alarm Fault Inhib...

Page 19: ...Off Default Off Hidden Area Up Down 0 9 Up Down Long Focus Up Long Focus Down Long Focus Menu Select Gas Range 0 Gas Range 100ppm 2 00 2ppm 0 Default 0 00 RS485 Addr Up Down Up Down Long Menu Select...

Page 20: ...50 0 ppm Alarm 2 Up Down 0 9 Up Down Long Focus Up Long Focus Down Long Focus Menu Select Alarm Level 2 0 9999 9 ppm Default 100 0 ppm Alarm Delay Up Down 0 9 Up Down Long Focus Up Long Focus Down Lon...

Page 21: ...rk Gateway 5 Time 6 Backlight Alarm Fault Backlight DHCP Up Down DHCP Menu Select On Network IP Address Off Network IP Address Default Off IP Address Up Down 0 9 Up Down Long Focus Up Long Focus Down...

Page 22: ...Up Down Long Focus Up Long Focus Down Long Focus Menu Select Date 2000 1 1 2099 12 31 Time 00 00 23 59 Backlight Up Down DHCP Menu Select Single Alarm Fault White Backlight Multi Alarm 1 Green Alarm...

Page 23: ...Span CAL Start Menu Span Calibration CAL Duration Span Calibration Span Gas Up Down 0 9 Up Down Long Focus Up Long Focus Down Long Focus Menu Select Span Gas 0 999 9 ppm Default 10 ppm CAL Duration Up...

Page 24: ...p Down Long Menu Select Flow 600 cc min Default 80 Hz 4mA CAL Up Down 0 9 Up Down Long Focus Up Long Focus Down Long Focus Menu Select 4mA Calibration 0 9999 4mA Default 1000 20mA CAL Up Down 0 9 Up D...

Page 25: ...Event Log Event Log List 10 7 Zero ADC Zero Calibration ADC 8 Span ADC Span Calibration ADC Up Down List Menu Menu Type Ex 03 O3 Event Type R Power On Alarm Fault Reset A Alarm F Fault I Information...

Page 26: ...3 MODBUS RS485 TCP Register 3000X Register Read Address Bits Description 30001 BIT15 0 Decimal Point Gas Range 30002 BIT15 0 Gas Range Decimal Point 1 Alarm 30003 BIT15 0 1 Alarm Decimal Point 2 Alarm...

Page 27: ...arm 1 Relay energized BIT9 Alarm 2 Relay energized BIT10 Fault Relay energized BIT11 Heartbeat Bit 2 Toggle BIT12 15 Reserved Cartridge Selection 40002 BIT0 7 Gas ID Sensor Type BIT8 15 Reserved 40003...

Page 28: ...Number 5 10 BIT8 15 Detector Serial Number 6 10 40034 BIT0 7 Detector Serial Number 7 10 BIT8 15 Detector Serial Number 8 10 40035 BIT0 7 Detector Serial Number 9 10 BIT8 15 Detector Serial Number 10...

Page 29: ...ower Loop Voltage Drop Power Supply voltage min operating voltage IMAX SI H100 WIRER m The resistance of the wire ohms meter value available in wire manufacturer s specification data sheet Ex 24V 16AW...

Page 30: ...mware Update 4 B 3 EEPROM Read Write MAIN Board 5 B 4 RTC Access MAIN Board 6 B 5 Reserved 7 Y 0 Reserved 8 S 0 Smart Sensor Smart Sensor 9 S 1 Smart Sensor Data Smart Sensor 10 S 2 Smart Sensor Expir...

Page 31: ...ial to Senko 14 Warranty 24 1 A S 18111 15 73 485 Tel 031 492 0445 Fax 031 492 0446 Email senko senko co kr www senko co kr 51610 275 2 1124 13 Tel 070 4633 3545 15 No Revision 1 Rev 1 0 2018 10 30 2...

Page 32: ...Operating Manual for Inhaled Gas Detector 31 Senko 2019 All Rights Reserved Proprietary and Confidential to Senko...

Reviews: