background image

ENGLISH

- 3/8

t

Power supply

RS485

MODBUS

Communicat.

1500 V

A

12

11

10

9

8

7

5

6

2

4
3

1

Input

TC

IDC10

IDC10

Input

TC

Input

TC

Input

TC

16

15

14

13

STANDARDS

EN61000-6-4

Electromagnetic emission, industrial environment
Electromagnetic immunity,  industrial environment

EN61000-6-2

Safety

EN61010-1

INSULATION

ENVIRONMENTAL COND.

Temp ratur

e

e

Hu

y

midit

Altitude
S rage temperature

to

P

ction degree

rote

-20 – +65°C.

EEPROM storing guaranteed in the range: 0    50 °C.

30% –

t condensing

90% no

Up to 2000 m a.s.l.
-20    + 85°C

20

IP

MOUNTING

IEC EN60715 DIN Rail

CONNECTIONS

Removable   -way terminals

4

,       mm

3.5

pitch, for cable max. 1.5mm .

2

Rear IDC10 connector for CEI EN60715 rail

Micro USB frontal panel connector.

POWER SUPPLY

Tension
Power absorbed

10 – 40 V

or 19 – 28 V

dc

ac

50    60 Hz

Max. 0.6W

COMMUNICATION
PORTS

RS485 on IDC10 rear connector.

PROTOCOLS

MODBUS-RTU.

TC INPUTS

Number of channels

Thermocouple

type: J, K, E, N, S, R, B, T.

8

TECHNICAL SPECIFICATIONS

LED

State

LEDs Meaning

PWR

(Green)

On

Power supply presence

Off

The device is powered off

FAIL

(Yellow)

On

Fault or Failure: power supply lack, failed channel, failed
T

, internal communication error.

hermocouple

RX (Red)

On

Data reception from

485 communication port.

RS

TX (Red)

On

Data transmission to

485 communication port.

RS

LED SIGNALLING ON FRONT PANEL

Summary of Contents for Z-8TC-LAB

Page 1: ...8 ThermoCouple Inputs ENGLISH 8 1 THE ORIGINAL VERSION IS IN ITALIAN LANGUAGE This document is property of SENECA rl Duplication and reproduction are forbidden if not authorized S Contents of the pres...

Page 2: ...rial communication with MODBUS RTU protocol l Frontal USB Port for MODBUS RTU con guration and communication l Channels individually activabl and con gurable in pairs e For the two inputs of each term...

Page 3: ...l CONNECTIONS Removable way terminals 4 mm 3 5 pitch for cable max 1 5mm 2 Rear IDC10 connector for CEI EN60715 rail Micro USB frontal panel connector POWER SUPPLY Tension Power absorbed 10 40 V or 19...

Page 4: ...sources PRELIMINARY WARNINGS TC type Allowed Range Linearization error TC type Allowed Range Lineariz error J 210 1200 C 0 05 C S 50 1768 C 0 02 C K 200 1372 C 0 05 C R 50 1768 C 0 02 C E 200 1000 C 0...

Page 5: ...ide of the IDC10 rear connector as shown in Pic 1 Removal from IEC EN 60715 DIN rail As shown in Pic 2 1 Move outwards the two hooks on the side of the module with the help of a screwdriver 2 Extract...

Page 6: ...e module will be inserted on a free slot of Z PC DIN accessory In the figure you can see the meaning of the various pins of the rear IDC10 connector if you want to provide the signals directly through...

Page 7: ...the Internet site www seneca com Some parameters can also be set using DIP switches CONFIGURATIONS Detailed information about list of MODBUS registers and their functions are available inside the Z 8...

Page 8: ...on Rejection ADC Filter Thermocouple type Both channels are enabled C Enabled 50 Hz ADC 15 bit with average filter J for both channels FACTORY SETTINGS DIP SWITCHES CONFIGURATION BAUD RATE SW1 1 2 960...

Reviews: