INSTRUCTIONS MANUAL CODE 80403 REV 0
09/2020 Pag.: 11
J
.P. SELECTA s.a.u.
Autovía A2 Km 585.1 Abrera 08630 (Barcelona) España
Tel 34 937 700 877 Fax 34 937 702 362
e-mail: [email protected] - website: www.jpselecta.es
EQUIPMENT DESCRIPTION
BLOOD-BANK A
storage cabinets are designed to work at fixed
temperatures at +4ºC (Stability ±1,5ºC). Equipped with an hermetic
compressor mounted on shock absorbers with evaporator and forced
air condenser, air cooling system which stops on opening the door,
temperature sensors housed in similar conditions to the blood bag,
inner LED light, double insulated anti-mist glass door and a circular
chart recorder.
External cabinet in white enamelled steel and interior in painted
aluminum and AISI 304 stainless steel base. Double glazed antimist,
reversible lockable door, easy to change, self close door it left ajar.
Fan air circulation in order to reach homogeneous temperature
distribution. External alarm ability. Circular graphic register. Pt100
probe. Safety thermostat. Temperature controlled by microprocessor
which thought display allows view and control following parameters:
- Security password
- Set temperature
- Set alarms
- Calibration
- Start up cooling motor delay.
Part
number
Litres
capac.
Bags
capac.
Internal dimensions cm
External dimensions cm
Consum
(W)
Power
(HP)
Nº
shelves
Weight
(Kg)
Height
Length
Width
Height
Length
Width
2101519
235
80
109
51
55
156
61
62
300
1/7
4
90
INSTALLATION
Avoid installing the equipment near heat sources (sunny places,
heating, etc). Leave a space of about 10 cm between the wall and
the back of the cabinet. Place the equipment on a hard horizontal
leveled floor; if it is necessary use the adjustable foot. Do not start
up before 4 hours.
VERY IMPORTANT BEFORE THE FIRST STARTING UP
Connect the equipment at mains.
The default temperature set is 4ºC. Do not fill the equipment if it has
not reached this temperature.