INSTRUCTIONS MANUAL
CODE 80382 REV B
02/2021 Pag.: 9
J
.P. SELECTA s.a.u.
Autovía A2 Km 585.1 Abrera 08630 (Barcelona) España
Tel 34 937 700 877 Fax 34 937 702 362
e-mail: [email protected] - website: www.grupo-selecta.com
EQUIPMENT DESCRIPTION
BLOOD-BANK C/D
storage cabinets are designed to work at fixed
temperatures at +4ºC (Stability ±1,5ºC) and
BLOOD-BANK E/F
platelet
incubators are designed for +22°C fixed temperatures. Equipped with
an hermetic compressor mounted on shock absorbers with evaporator
and forced air condenser, air cooling system which stops on opening
the door, temperature sensors housed in similar conditions to the blood
bag, inner LED light, double insulated anti-mist glass door and a circular
chart recorder.
Stainless steel AISI 304 external cabinet. Inner lining also in stainless
steel. Double glazed antimist, reversible lockable door, easy to change,
self close door it left ajar.
Fan air circulation in order to reach homogeneous temperature
distribution. External alarm ability. Circular graphic register. Pt100 probe.
Safety thermostat. Temperature controlled by microprocessor which
thought display allows view and control following parameters:
- Security password
- Set temperature
- Set alarms
- Calibration
- Start up cooling motor delay.
Part
number
Litres
capac.
Bags
capac.
Internal dimensions cm
External dimensions cm
Consum
(W)
Power
(HP)
Nº
shelves
Weight
(Kg)
Height
Length
Width
Height
Length
Width
2101512
600
200
136
58
69
207
70
82
350
3/8
5
132
2101514
1300
480
136
128
69
207
140
82
1100
3/8
10
190
2101511
600
200
136
58
69
207
70
82
350
3/8
5
132
2101513
1300
480
136
128
69
207
140
82
1100
3/8
10
190
INSTALLATION
Avoid installing the equipment near heat sources (sunny places, heating,
etc). Leave a space of about 10 cm between the wall and the back of
the cabinet. Place the equipment on a hard horizontal leveled floor; if
it is necessary use the adjustable foot. Do not start up before 4 hours.
VERY IMPORTANT BEFORE THE FIRST STARTING UP
Connect the equipment at mains.
The default temperature set is 4ºC (22ºC in the case of models E/F). Do
not fill the equipment if it has not reached this temperature.