13 Controller configuration
1. Power switch
7. Condenser
2. Microcomputer controller
8. Drain case
3. Compressor
9. Shelf
4.
Evaporator
10.
Lamp
5. Evaporator fan motor
6. Condenser fan motor
1. New upright air-cooling refrigerator should be first opened and ventilated before
its use.
After that, users should use warm water to clean inside.
2. After connecting the power supply, press the “POWER” switch on the controller
keyboard (Green Indicator Light ON), the fridge will come to work. The microcom
-
puter controller, installed in the controller keyboard, could automatically adjust the
temperature ranges. This intelligent digital controller works as: if the temperature
increases and reaches set point plus differential the compressor is started and then
turned off when the temperature reaches the set point value again.
3.
To display target set point, select a parameter in programming mode or
confirm an operation.
(DEF) To start a manual defrost.
(UP) In programming mode, browse parameter codes or increase the
displayed value.
(DOWN) In programming mode, browse the parameter codes or decrease
the displayed value.
To lock or unlock the keyboard.
To enter in programming mode.
To return to room temperature display.
4. How to see the set point
Press and immediately release the SET key, the set point will be showed;
Press and immediately release the SET key or wait about 5s to return to normal
visualisation.
5. How to change the set point
Press the SET key for more than 2 seconds to change the set point value;
The set point value will be displayed and the “ºC” or “ºF” LED starts blinking;
To change the set value, press UP or DOWN arrows.
To memorise the new set point value, press SET key again or wait 10s.