Requirements according to IEC 60601-1-2:2007
OLA 8 & 4 are intended for use in the electromagnetic environment specified
below.
Guidance and manufacturer’s declaration –
electromagnetic emissions
Emission test
Compliance
Electromagnetic
environment – guidance
RF emission
CISPR 11
Group 1
The control unit uses RF energy only for
its internal function. Therefore, its RF
emissions are very low and are not likely
to cause any interference in nearby
electronic equipment.
RF emission
CISPR 11
Class B
OLA 8 & 4 are suitable for use in all
establishments, including domestic
establishments and those directly
connected to the public, low-voltage
power supply network that supplies
buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
25
•
The control unit should not be used next to or stacked with
other equipment where possible. If this is unavoidable the
control unit should be observed to verify normal operation.
Select Medical Ltd: 17 April 2018
OLA 8 & 4 UM20180417 – Review V1.1