OXIGEN SENSOR (O
2
) CALIBRATION DETAIL
Apply gas to the instrument and adjust the output pressure of the gas from the cylinder so that the flow meter
indicates a minimum flow of 0.5 l/m: this guarantees that the instrument is taking the exact amount of gas
required by the internal pump.
The instrument measures the concentration of gas applied;
wait at least 3 minutes to allow the reading to
stabilize
. The reading is shown in line 'Meas'.
The calibration
is possible
only when the status is set to ‘
----
' (sensors that have never been calibrated
before) otherwise it is necessary to set 'Status' to '
OFF
' (see example below).
11/12/17 10:00
ON SITE CALIB.
Calibration
►
Status ----
Durat. 00:00:00
Appl
0.0 %
Meas
21.0 %
Is 99.650 uA
or
>
>
11/12/17 10:00
ON SITE CALIB.
Calibration
►
Status ON
Durat. 00:00:02
Appl
0.0 %
Meas
21.0 %
Is 99.650 uA
OK
11/12/17 10:00
ON SITE CALIB.
Calibration
►
Status
Durat. 00:00:03
Appl
0.0 %
Meas
21.0 %
Is 99.650 uA
ON
11/12/17 10:00
ON SITE CALIB.
Calibration
►
Status
Durat. 00:00:04
Appl
0.0 %
Meas
21.0 %
Is 99.650 uA
OFF
OK
11/12/17 10:00
ON SITE CALIB.
Calibration
Status ----
►
Durat. 01:03:15
Appl
0.0 %
Meas
21.0 %
Is 99.650 uA
Resets the timer (helps to control
the elapsed time during the
stabilization phase)
OK
11/12/17 10:00
ON SITE CALIB.
Calibration
►
Status ----
Durat. 01:30:11
Appl
0.0 %
Meas
21.0 %
Is 99.650 uA
>
11/12/17 10:00
ON SITE CALIB.
Calibration
Status ----
►
Durat. 00:00:00
Appl
0.0 %
Meas
21.0 %
Is 99.650 uA