4. Operation
Controls and displays
How to weigh properly
– Press the
start
key with no load on the
scale.
SE(A
,
8.8.8.8.8.
and
0.00
appear con-
secutively in the display. The scale is
then automatically set to zero and ready
for use.
– If necessary, switch the weight display
(see page 15).
– Ask the patient to mount the scale.
Once on the scale, the patient should
stand still.
The HOLD function is activated auto-
matically at weights of over 1.5 kg or
3.3 lbs. The display flashes until a stable
weight has been measured. The display
is then frozen until the next weighing op-
eration.
HOLD
appears in the
display.
– Read off the weighing result on the
digital display.
– If a load greater than 200 kg or 440 lbs
is placed on the scale,
St0P
appears in
the display.
– To switch off the scale, press the
start
key again.
start
key
ON-OFF switch
2 in 1
key/
to switch between kg and lbs
kg
61.40
HOLD
kg
St0p