7.0
Electrical Specification
Appendix 1 specifies the operating error for all tests and functions of the
PrimeTest 200.
7.1
Earth continuity measuring range (rotary switch position , and IEC)
Measuring Range
Resolution
Accuracy
Overload Protection
19.99
Ω
0.01
Ω
±5% +2 digits
300 V AC/DC
Analogue bargraph display
0 to 20
Ω
Bargraph accuracy
10%
Open circuit voltage
>4VDC
Test current
>200mA (0-2
Ω
)
Input protection
0.5A (600V) quick blow fuse
An input voltage of
≥
30 V AC/DC, detected between the + and - input measuring
terminals, disables the earth continuity measuring circuit. A repetitive audible
alarm is activated and a flashing high voltage symbol is displayed on the LCD.
The maximum number of continuity tests, with a duty cycle specified by
BS EN 61557-4 Section 6.6, is 1900 repetitions.
7.2
Insulation resistance measuring range (rotary switch position , , IEC
and M
Ω
.
Measuring Range
Resolution
Accuracy
Overload Protection
2M
Ω
0.01M
Ω
±2% +2 digits
300 V AC/DC
20M
Ω
0.1M
Ω
±2% +2 digits
300 V AC/DC
200M
Ω
1M
Ω
±5% +2 digits
300 V AC/DC
Analogue bargraph display
0 to 200M
Ω
and infinity
Bargraph accuracy
10%
Test voltage
>500VDC at 1mA, <750VDC for o/c
Test voltage accuracy
+20%, -0%
Test current
<2mA for s/c
Input protection
0.5A (600 V) quick blow fuse.
An input voltage of
≥
30 V AC/DC, detected between the + and - input measuring
terminals, disables the insulation resistance measuring circuit. A repetitive
audible alarm is activated and a flashing high voltage symbol is displayed on
the LCD.
The maximum number of insulation tests, with a duty cycle specified by
BS EN 61557-2 Section 6.7, is 2000 repetitions.
7.3
Resistance measuring range
Measuring Range
Resolution
Accuracy
Overload Protection
19.99
Ω
0.01
Ω
±5% +2 digits
300 V AC/DC
1999
Ω
1
Ω
±5% +2 digits
300 V AC/DC
20.00K
Ω
10
Ω
±5% +2 digits
300 V AC/DC
Analogue bargraph display
0 to 20K
Ω
and infinity
Bargraph accuracy
10%
Test current
>200mA (0-2R)
Open circuit voltage
> 4 V DC
Input protection
0.5A (600 V) quick blow fuse.
Zero compensation of the test leads up to 10
Ω
. The internal buzzer sounds
when the measured resistance <10
Ω
.
An input voltage of
≥
30 V AC/DC, detected between the + and - input measuring
terminals, disables the resistance measuring circuit. A repetitive audible alarm is
activated and a flashing high voltage symbol is displayed on the LCD.
PrimeTest 200
Operating Instructions
5
Summary of Contents for PrimeTest 200
Page 3: ...Operating Instructions PrimeTest 200 Fig 5 Fig 6...
Page 4: ...Operating Instructions PrimeTest 200 Fig 7 Fig 8...
Page 5: ...Operating Instructions PrimeTest 200 Fig 9 Fig 10...
Page 6: ...Operating Instructions PrimeTest 200 Fig 11...
Page 7: ...Operating Instructions PrimeTest 200 Fig 12B Fig 12A...
Page 8: ...PrimeTest 200 Operating Instructions...
Page 10: ...PrimeTest 200 Operating Instructions...