Seametrics EX100 Series Instructions Manual Download Page 6

EX100/200 SERIES INSTRUCTIONS

Seametrics • 253.872.0284 

Page 6 

seametrics.com

Fitting Installation. 

 EX11x/21x adapters mate with a 1-1/2” female 

NPT  pipe  thread  adapter  fitting.    Any  fitting  that  provides  the 

matching NPT female thread may be used.  Installation procedure 

compensates for fitting height differences. Cut a minimum 1-3/4” 

hole in the pipe.  If possible, measure the wall thickness and write 

it down for use in depth setting.  Then install the threaded fitting 

(saddle, weldolet, etc.) on the pipe.

Meter Installation. 

Loosen the compression nut so that the adapter 

slides  freely.    Pull  the  meter  fully  upward  and  finger-tighten  the 

compression nut.  Using a thread sealant, install the adapter in the 

pipe fitting.  Do not overtighten.  Now loosen the compression nut, 

lower the meter to the appropriate depth setting (see diagram and 

instructions that follow).  Be sure flow is in the direction of the arrow 

on the housing.  Fully tighten compression nut.  

‘Hot tap’ EX meters are designed so they can be installed and 

serviced without depressurizing the pipe.

Fitting Installation. 

 The EX15x/25x adapters mate with a 2” FNPT 

threaded  fitting  for  compatibility  with  the  2”  isolation  valve.    Any 

fitting that provides matching NPT female thread may be used.  The 

installation procedure compensates for differences in fitting height. 

If  initial  installation  is  performed  on  an  unpressurized  pipe,  cut  a 

minimum 1-3/4” hole in the pipe.  If possible, measure the wall 

thickness and write it down for use in depth setting.  Then install the 

threaded fitting (saddle, weldolet, etc.) on the pipe. 

If it is necessary to do the initial installation under pressure, any 

standard hot tap drilling machine with 2” NPT adapter, such as a 

Transmate or a Mueller, can be used.  Ordinarily, it is not necessary 

to use an installation tool, due to the small diameter tube the meter 

can be installed by hand at all but the highest pressures. 

Meter Installation. 

 Remove the sensor unit from the valve assembly.  

Using a thread sealant, install the valve assembly on the pipe fitting.  

If the initial installation is a pressure (“hot”) tap, remove the 1-1/2” 

x 2” adapter bushing at the back of the valve.  Thread the tapping 

machine on, open the valve, and tap using a minimum of 1-3/4” or 

maximum 1-7/8” cutter.  After retracting the machine and closing 

the valve, reinstall the flow sensor.  When the sensor is secure, open 

the valve and adjust depth setting (see diagram and instructions that 

follow). Be sure flow is in the direction of the arrow on the housing. 

Fully tighten locking collar and compression nut.

INSTALLATION

Compression nut

Adapter mates with

FNPT threads

EX15x/25x Sensor 

Removal

Compression 

nut

2” adapter removes 

to mount hot-tap 

machine

Full-port 2” ball 

valve allows sensor 

removal

Mates with 2” 

FNPT threads

Locking collar

EX11x/21x Installation

EX15x/25x Installation

Summary of Contents for EX100 Series

Page 1: ...PROUDLY MADE IN THE USA C e rtified Compa n y ISO 9001 2008 EX100 200 Series Insertion Magnetic Flow Meter Instructions EX11x EX21x EX15x EX25x...

Page 2: ...ections General Electrical Guidelines Page 11 Power Page 11 Output Page 11 Grounding Guidelines Page 11 Connection Diagrams Counter or PLC Page 12 A055 4 20 mA Output Page 12 FT520 Controller Page 12...

Page 3: ...es converted using any of the Seametrics family of indicators and converters A modular system of electronics can be installed directly on the flow sensor or mounted remotely The FT430 externally power...

Page 4: ...Valve Assembly 15x 25x only Bronze stainless optional with bronze ball valve Fitting Size Required 11x 21x 1 5 FNPT 15x 25x 2 FNPT Maximum Pressure 200 psi 14 bar Temperature Ambient 0 to 160 F 17 to...

Page 5: ...are not recommended for installation downstream of a boiler feedwater pump where installation fault may expose the flow sensor to boiler pressure and temperature Maximum recommended temperature is 20...

Page 6: ...ht If initial installation is performed on an unpressurized pipe cut a minimum 1 3 4 hole in the pipe If possible measure the wall thickness and write it down for use in depth setting Then install the...

Page 7: ...ension D one more time 5 Fully tighten the compression nut Record your settings Once you have the meter set up and operational it is important to record you meter settings and save them future referen...

Page 8: ...37 1 031 1 218 Stainless Steel 10S 0 120 0 120 0 134 0 148 0 165 0 180 0 188 0 188 0 188 0 218 0 250 0 312 0 312 Stainless Steel 40S 0 216 0 237 0 280 0 322 0 365 0 375 0 375 0 375 0 375 0 375 0 375 0...

Page 9: ...age 9 seametrics com INSTALLATION Straight Pipe Recommendations X diameter 5X 10X Reduced Pipe Two Elbows In Plane Two Elbows Out Of Plane Expanded Pipe Swirling Flow Propeller Meter Partially Open Bu...

Page 10: ...ault may expose the flow sensor to boiler pressure and temperature Maximum recommended temperature is 200 F Full Pipe Recommendations Better Installation Possible Problem Air can be trapped Allows air...

Page 11: ...cable be sure to connect shield to ground at power supply end of the cable Do not connect other end of shield Avoid routing flow sensor cables in close proximity to a variable frequency drive Recomme...

Page 12: ...dc Power Forward Output EX SERIES AO55 EX SERIES Sensor 4 20 mA 4 20 mA 24 Vdc Power 12V SEN1 G SEN2 EX SERIES 24 Vdc Power Power Forward Output FT520 Reverse Output Option 15 only Reverse Output Opti...

Page 13: ...See Dual FT430 440 Diagram for an example of bidirectional connections _ _ _ _ _ _ _ S _ _ _ Power Pulse Out 2 Sensor Input Pulse Out 1 Scaled Scaled Reverse Output Option 15 only Power Forward Output...

Page 14: ...ions Green White Black Red 12 24 Vdc Power _ S _ _ _ _ _ S _ _ _ Power Pulse Out 2 Sensor Input Pulse Out 1 Scaled 430 Pass Thru 440 Scaled 24 Vdc Power DL76 FT430 440 EX SERIES Sensor Input Reverse O...

Page 15: ...es it may be necessary to adjust the flow meter under no flow conditions after it has been installed This should only be done if the indicated flow is low near the lower cutoff Zero Adjustment To perf...

Page 16: ...h 100414 100414 7 Strain relief kit large includes 1 101850 101850 9 Tube Not field replaceable 10 Compression nut 100064 brass 100084 ss 11 Compression ferrule 100358 12 Adapter 100845 brass 100846 s...

Page 17: ...ons Fluid conductivity 20 S cm Select another flow meter Output pulses incorrect Missing or incorrect ground wire Check for proper ground Incorrect depth setting Check depth setting see page 7 Fluid c...

Page 18: ...Seametrics which cause damage to the products or cause abnormally frequent service calls or service problems h A new warranty period shall not be established for repaired or replaced material products...

Page 19: ...EX100 200 SERIES INSTRUCTIONS Seametrics 253 872 0284 Page 19 seametrics com NOTES...

Page 20: ...Seametrics 19026 72nd Avenue South Kent Washington 98032 USA P 253 872 0284 F 253 872 0285 1 800 975 8153 seametrics com LT 65200196r2 1 20161230 12 30 16...

Reviews: