5. MAINTENANCE
5.1.
PAPR
CLEANING:
5.1.1.
Detach the battery pack, hose assembly and the blower. Inspect all parts for damage. Replace all damaged parts prior to storage
or next use.
5.2.
BLOWER
5.2.1.
Clean the outer surfaces of PAPR assembly and battery pack with a soft cloth dampened in a solution of water and mild, pH
neutral
detergent.
DO NOT
immerse the blower or battery pack in water.
DO NOT
use solvents or abrasive cleaners.
DO NOT
attempt to clean the interior of the blower with compressed air or vacuum.
Ensure the electrical contacts of the blower and battery pack are dry.
5.3.
HOSE ASSEMBLY
5.3.1.
Clean the connection joints on the hose assembly with water and a mild detergent solution.
NOTE:
The hose assembly can be immersed in water for cleaning.
5.3.2.
Air dry or dry by connecting to the blower unit and blow dry the hose.
NOTE:
The inside of the hose must be completely dried prior to use or storage.
5.4.
HEPA FILTER
5.4.1.
Remove
the
fi
lter cover and inspect the
fi
lter.
5.4.2.
Replace if excessively dirty.
5.5.
CHANGING THE VISOR
5.5.1.
Release the stud from the bottom of the outside front of the face shield.
5.5.2.
Release the two tags on the outside bottom edge of the visor.
5.5.3.
Release the two sides and pull downwards to release the visor.
5.5.4.
Fitting a new visor is the reverse of the above.
5.6.
CLEANING AND DISINFECTION THE VISOR
DO
NOT
use abrasive materials.
5.6.1.
The lens is not scratch resistant. Be very careful when wiping, keep this to a minimum and use a soft, non-friction cloth.
5.6.2.
Clean with a soap and warm water or with ordinary household disinfectant.
5.7.
INSPECTION
5.7.1.
Face shields should be periodically inspected and cleaned. If it is damaged , pitted or anything that seriously a
ff
ects the vision
through it, the lens must be replaced.
5.8.
STORAGE
5.8.1.
If the face shield with powered air purifying respirator is not to be used for an extended period, remove the
fi
lter and battery pack
and store in its original packaging in a clean, dry, cool place free of solvent based vapours.
5.8.2.
Face shield: Good storage conditions are very important for the service life of the face shield. It should be stored at temperatures
between 5ºC to 40ºC and 90% relative humidity. It should be kept away from solvents, solvent vapours or any corrosive materials,
because the protection e
ff
ect of the face shield may be seriously reduced.
5.8.2.1.
When not in use it is recommended that the face shield is stored in its original packaging in a clean, dry, cool place.
NOTE:
Safety visor screen (SSP80PAPR.13) only, has a product life of 2 years from date of purchase.
6. TROUBLESHOOTING
FAULT
POSSIBLE SOLUTION
Blower does not supply air to hood
1. Press ON/OFF to start blower
2. Recharge battery
3. Verify battery is properly connected to blower
4. Remove blockage from blower outlet and breathing tube
Battery pack’s charge lasts less than expected
1. Ensure battery pack is fully charged
2. Replace battery
3. Replace charger
4. Check the air
fi
lter (HEPA
fi
lter & Foam pre
fi
lter), and replace it if necessary
Blower cannot be turned o
ff
Press ON/OFF button for one second
Battery red level light is
fl
ashing
Charge or replace the battery
Danger light is on, alarm sounds or blower vibrates
Check the blower air
fl
ow as section 4.11
SSP80PAPR Issue 1 11/08/21
Original Language Version
© Jack Sealey Limited