Cleaning the Prep Station
1. Fill up two 50 mL conical tubes with DI water. Fill up one 50 mL conical tube with 70% ethanol.
2. Remove the cap containing the tube from the assay medium bottle and place the tube into one
of the 50 mL conical tube containing DI water from #1.
3. Click “Prime Manifold” from the tab of the prep station software.
4. Select “Media”
5. Click “Start”
6. After the cycle is finished, transfer the tube to the 50 mL conical tube containing 70% ethanol.
7. Click “Start”
8. After the cycle is finished, transfer the tube to the second 50 mL conical tube containing DI
water.
9. Click Start.
10. After the cycle is finished, put the cap containing the tube back into the empty media bottle.